47
Subject to change without notice
C a l i b r a t i o n
6 Calibration
Prior to any measurement an open circuit, resp. a short circuit
calibration should be performed in order to avoid measurement
errors. The open/short circuit calibration compensates for the
effects of parasitic impedances of the connections to the compo-
nent under test. A load calibration is especially suitable in order
to take an impedance measured before the measurement of the
unknown into account. With a calibration it is possible to com-
pensate measurement connections and other parasitic effects
(capacitive impedances). Any calibration must be performed at
the actual measuring frequency.
In order to achieve the best accuracy all three calibration steps
should be carried out in the same set-up and the same environ-
mental conditions used for the measurement of the unknown,
especially the placement of the leads should remain identical.
The measurement leads should be far from any metallic objects
and the hands of the operator. In order to initiate the calibration
steps the respective parameter must be set to ON by pressing
SELECT
3
and calling the menu CORR. The parameters OPEN /
SHORT / LOAD can now be set using the arrow buttons
8
Änderungen vorbehalten
B e z e i c h n u n g d e r B e d i e n e l e m e n t e
2 Bezeichnung der Bedienelemente
Gerätevorderseite
1
POWER (Taste)
Netzschalter; Netzanschluss auf der Geräterückseite
2
GATE (LED)
Die GATE-LED leuchtet während der gesamten Dauer einer
Messung. Dies entspricht der gewählten Torzeit und einer
Synchronisierungszeit.
3
REMOTE (LED und Taste)
Die REMOTE–LED leuchtet, sobald das Gerät über die Schnitt-
stelle angesprochen wird. Um zur manuellen Betriebsart
zurückzukehren, ist die REMOTE-Taste zu drücken.
4
Display (LCD-Anzeige)
Anzeige des Messergebnisses und verschiedener Zusatzin-
formationen
5
ESC (Taste)
Escape-Taste in der Menüsteuerung
6
ENTER (Taste)
Enter-Taste in der Menüsteuerung
7
SELECT (Taste)
Menüaufruf bzw. Auswahl eines Menüpunkts
8
(Tasten)
Pfeiltasten zur Menüsteuerung und Parametereinstellung
9
Drehgeber
Drehknopf zur Parametereinstellung
10
GATE TIME (Taste)
Einstellung der GATE-Zeit
11
LEVEL B (Taste)
Einstellung des Triggerlevels von Kanal B
12
LEVEL A (Taste)
Einstellung des Triggerlevels von Kanal A
13
16
1 : 10 (Taste)
Eingangssignalabschwächer, Gesamtabschwächung 100-fach
14
DC (Taste)
Wahl der Kopplungsart des entsprechenden Kanals:
Taste DC leuchtet = DC-Kopplung
Taste DC aus
= AC-Kopplung
15
Slope (Taste)
Durch Drücken dieser Taste wird die Triggerflanke gewählt.
Leuchtet die Taste, wird auf die negative Flanke getriggert. Ist
die Taste unbeleuchtet, erfolgt die Triggerung auf die positive
Flanke.
17
50 Ω (Taste)
Zuschalten eines 50 Ω-Widerstands zum Eingang zur Anpas
-
sung bei 50 Ω-Systemen
18
LP 50 kHz (Taste)
Tiefpassfilter zur Vermeidung unerwünschter HF-Triggerung
bei niederfrequenten Signalen
19
23
TRIG (LEDs)
Triggerindikatoren
20
22
INPUT A, INPUT B (BNC–Buchsen)
Messsignaleingänge DC-200 MHz
21
AUTO TRIG (Taste)
Aktivierung des Auto-Triggers. Die Taste AUTO TRIG leuchtet,
wenn die automatische Triggerung aktiv ist.
24
INPUT C (SMA-Buchse)
Messsignaleingang 100 MHz – 3 GHz
25
RESET · V
Taste mit Doppelfunktion:
1. Durch Drücken dieser Taste wird die laufende Messung
unterbrochen, die Anzeige gelöscht und die Messung neu
gestartet.
2. Bei Einstellung des Triggerlevels mit den Zifferntasten wird
der eingegebene Wert mit der Einheit Volt (V) übernommen.
26
TRIG · GHz/s (Taste)
Taste mit Doppelfunktion:
1. Auslösen einer Messung im ARMED-Betrieb.
2. Bei Einstellung der Gatetime mit den Zifferntasten wird der
eingegebene Wert mit der Einheit Sekunde (s) übernommen.
27
HOLD · mV (Taste)
Taste mit Doppelfunktion:
1. Durch Drücken dieser Taste wird der zuletzt im Display
angezeigte Messwert eingefroren.
1
2
3
4
5 6 7
8
9
10 11 12
13
14 15
16
17 18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
7
or with the knob
6
. The calibration procedure takes a few
seconds; as soon as it has been performed successfully, the
display of the selected calibration message will disappear.
13
12
11
Fig. 6.1: Buttons for the calibration
From the firmware version 1.35 upwards, the HM8118 offers
the possibility to automatically extend the OPEN and SHORT
calibration steps over all 69 available frequencies. Select the
menu item CORR and change the entry MODE from SGL (single)
to ALL. Then leave the menu und press the pushbuttons OPEN
resp. SHORT on the front panel.
Now the HM8118 will perform an automatic calibration of
OPEN and SHORT at all 69 frequencies. This procedure will
take appr. 90 seconds. After a successful calibration a short
beep will occur.
For further details on OPEN resp. SHORT calibration see section
6.1 resp. 6.2.
6.1 Open circuit calibration procedure
In order to perform a correct open circuit calibration there must
be nothing connected to the measurement leads! To start the
calibration function press the button OPEN
11
.
The display will show
„Opening: (measuring frequency)“
.
If no calibration can be performed successfully, an error mes-
sage will be displayed. The open circuit calibration is possible
if the impedance is
>
10 kΩ.
6.2 Short circuit calibration procedure
In order to perform a correct short circuit calibration the mea-
surement leads must be shorted directly with no components
attached. To start the calibration press the SHORT button
12
.
The display will show
„Shorting (measuring frequency)“
.
If no calibration can be performed successfully, an error mes-
sage will be displayed. The short circuit calibration is possible
if the impedance is
<
15 Ω or the resistance
<
10 Ω.
6.3 Procedure of the calibration with a known
load
After the selection of the desired measuring function, e.g.
L-Q
,
the calibration with a known load is performed separately for the
main display LOADM and the second display LOADS (S stands
for “sub”). It is recommended to choose the value of this load
close to the expected value of the DUT (= Device Under Test).
Press the button LOAD
13
to start the calibration.
The display will show
„Load Cal: [measuring frequency]“
.
If no calibration can be performed successfully, an error mes-
sage will be displayed. The LOAD calibration is possible for
impedances and resistances within the full measurement range.
Any calibration with a known load requires that
values were entered in both reference memories
(LOADM and LOADS). (In the above example a
value for the expected inductance in LOADM and a
value for the expected quality factor Q in LOADS.
The latter is only valid for the selected measuring
frequency.)
Summary of Contents for HM8118
Page 1: ...P r o g r a m m a b l e L C R B r i d g e H M 8 1 1 8 Handbuch Manual Deutsch English...
Page 57: ...57 Subject to change without notice C o m m a n d r e f e r e n c e...
Page 58: ...58 Subject to change without notice C o m m a n d r e f e r e n c e N o t e s N o z i z e n...
Page 59: ...59 Subject to change without notice C o m m a n d r e f e r e n c e N o t e s N o t i z e n...