background image

Use and Compensation of Probes

To 

display an undistorted waveform on an oscilloscope, the

probe must be matched to the individual input impedance
of the vertical amplifier.

The 

 built-in calibration generator provides a

squarewave signal with a very low 

 

 and

switch-selectable frequencies of approx. 

 and 1 

MHz

at two output sockets below the CRT 

screen. One output

provides 

 

 % 

for 

10: 1 

probes, and 

 

 % are

present at the other, for 100: 1 probes.

When the attenuator switches are set to 

 vertical

deflection coefficient, these calibration voltages corre-

spond to a screen amplitude of 

The output sockets have an internal diameter of 

 to

accommodate the internationally accepted shielding tube

diameter of modern 

 

 and 

F-series 

probes. Only this type of construction ensures the
extremely short ground connections which are essential for
an undistorted waveform reproduction of non-sinusoidal
high frequency signals.

Adjustment at 1 

The C-trimmer adjustment compensates the capacitive

loading on the oscilloscope input (approx. 

 with the

HM604). 

By this adjustment, the capacitive division

assumes the same division ratio as the ohmic voltage
divider to ensure an equal division ratio for high and low fre-
quencies, as for DC. (For 

1: 1 

probes or switchable probes

set to 1: 1, this adjustment is neither required nor possible).

A baseline exactly parallel to the horizontal graticule lines is

a major condition for accurate probe adjustments. (See also

‘Trace Rotation 

TR’, 

page M7.)

Connect the probes (Types HZ51, 52, 53, 

54, 

or HZ37) to

CH.1 

input. All pushbuttons should be released (in the ‘out’

position), and all push-pull knobs pushed ‘in’. Set the input

coupling switch to 

DC, 

the 

attenuator switch to 

and the 

 switch to 

 and all variable

controls to 

CAL. 

position. Plug the probe tip into the appro-

priate calibrator output socket, i.e. 

 probes into the

 socket, 100: 1 probes into the 

 socket.

incorrect

correct

incorrect

Approximately 2 complete waveform periods are displayed
on the CRT screen. Now the compensation trimmer has to

be adjusted. Normally, this trimmer is located in the probe
head. On the 100: 1 probe HZ53, 

however, it is located in

the connecting box at the other end of the cable. Using a

small insulated non-metallic screwdriver or trimming tool,

the trimmer has to be adjusted slowly until the tops of the

squarewave signal are exactly parallel to the horizontal

graticule lines. (See Fig. above for 

 The signal

amplitude shown should be 4cm 

 1.2 

mm 

 3 

 During

this adjustment, the signal edges will remain invisible.

Adjustment at 1 MHz

Probes 

 and 54 will also allow for HF-adjustments.

They incorporate resonance deemphasizing networks 
trimmer in conjunction with inductances and capacitors)
which permit   for the first time   probe compensation in
the range of the upper frequency limit of the vertical oscillo-
scope amplifier.

Only this compensative adjustment

ensures optimum utilisation of the full bandwidth, together
with constant group delay at the high frequency 

end,

thereby reducing characteristic transient-distortion near the

leading signal edge (e.g. overshoot, rounding, ringing, holes

or bumps) to an absolute minimum.

Using the probes 

 52, 

and 54, the full bandwidth of

the HM 604 can be utilized without risk of unwanted wave-
form distortion.

Prerequisite for this HF-adjustment is a squarewave

generator with fast 

 (typical 

 and low output

impedance (approx. 

 providing 

 and 2V at a fre-

quency of approx. 

MHz. The calibrator output of the

HM604 meets these requirements when the pushbutton

1 MHz is depressed.

Connect the probe 

 or 54) to CH.1 input. Depress

the calibrator pushbutton 

 All other pushbuttons

should be released (‘out’ position). Set the input coupling

switch to DC, attenuator switch to 

 and TIME/

DIV. switch to 0.1 

 Set all variable controls to CAL.

position.

Insert the probe tip into the output socket marked 

 A

waveform will be displayed on the CRT screen, with leading

and trailing edges clearly visible. For the HF-adjustment

now to be performed, it will be necessary to observe the ris-

ing edge as well as the upper left corner of the pulse top. To

gain access to the 

 trimmer, the plastic

cover of the probe connecting box has to be slid off after

unscrewing the probe cable. The connecting boxes of the

HZ51 and HZ54 contain one R-trimmer screw, each, while

that of the HZ52 provides three. These R-trimmers have to

be adjusted in 

such a manner that the beginning of the pulse

top is as straight as possible. Overshoot or excessive round-

ing are unacceptable. This is relatively easy on the HZ51

and HZ54, but slightly more difficult on the HZ52. The rising
edge should be as steep as possible, with the pulse top

remaining as straight and horizontal as possible.

On the HZ52, each of the three trimmers has a clearly

M8 604

Subject to change without notice

Summary of Contents for HM 604

Page 1: ...er Distortions T 1 Astigmatismus Check T 1 Symmetry and Drift of thevertical Amplifier T 1 Calibration of the Vertical Amplifier T 1 Transmission Performance of the Vertical Amplifier T 2 Operating Mo...

Page 2: ...1 kHz to 1 MHz t approx 3ns Output voltage 0 2V and 2V fl Protective system Safety Class I IEC 348 Linevoltage 110 125 220 24OV IO Line frequency 50 to 60Hz Power consumption 40 Watt Weight approx 8kg...

Page 3: ...MQ switchable Attenuation ratio 1000 1 Bandwidth 1 MHz Cable length 1 5 m BNC connector Test Cable Banana BNC HZ32 Coaxial test cable length 1 15 m characteristic impedance 5Ofi Cable capacitance 12Op...

Page 4: ...tact of the appliance inlet The instrument operates ac cording to Safety C ass I three conductor power cord with protective earthing conductor and a plug with earthing con tact The mains line plug sha...

Page 5: ...g fluid get into the instrument The use of other cleaning agents can attack the plastic and paint surfaces Switching over the mains line voltage The specifications stating tolerances are only valid if...

Page 6: ...voltage data normally refers to effective values rms root mean square value However for signal magnitudes and voltage designations in oscilloscope measurements the peak to peakvoltage V value is appl...

Page 7: ...t D 62 2 3 2 19 4V cm min deflection coefficient D 62 2 8 7 8V cm adjusted deflection coefficient D lOV cm If the applied signal is superimposed on a DC direct voltage level the total value DC peak va...

Page 8: ...r line Adjust the Y at tenuator switch with its variable control together with the Y POS control so that the pulse height is precisely aligned with the 0 and 100 lines The 10 and 90 points of the sign...

Page 9: ...the ground point on the test circuit can be critical This ground point should al ways be located as close as possible to the measuring point If this is not done serious signal deformation may result f...

Page 10: ...leased out position If the trace is not visible check the correct positions of all knobs and switches particularly LEVEL knob in AT position and DELAY MODE lever switch to OFF To obtain the maximum li...

Page 11: ...adjusted slowly until the tops of the squarewave signal are exactly parallel to the horizontal graticule lines See Fig above for 1 kHz The signal amplitude shown should be 4cm 1 2 mm 3 During this adj...

Page 12: ...rticular importance Providing these essential features as well as switch select able output frequencies the calibrator of the HM 604 can under certain conditions replace expensive squarewave generator...

Page 13: ...e fluorescent screen of the CRT For X Yoperation the pushbutton in the X Section marked X Y must be depressed The X signal is then derived from the Channe HOR INP The calibration ofthexsignal during X...

Page 14: ...ude modulated oscillation F 1 MHz f 1 kHz m 50 UT 28 3 mVrms Oscilloscope setting for a signal according to figure 2 Depress no buttons Y CH I 20mV div AC TIM E DIV 0 2 ms div Triggering NORMAL with L...

Page 15: ...s a display below the internal trigger threshold Line triggering is recommended for all signals which are time re lated multiple or submultiple to the mains line frequency or when it is desirable to p...

Page 16: ...again after delay provided there is another suitable edge DEL TRIG In DEL TRIG mode andN SEP switch in H or V position after delay triggering to the next following line is possible Therefore discrete...

Page 17: ...increased the trace will tend to shift If this happens the VAR delay time control can be readjusted also sub sequently at any time to enable the exact point of interest to be displayed In the example...

Page 18: ...d without any restriction Using the Component Tester The CT is switched on by depressing the COMPONENT TESTER pushbutton This makes the vertical preamplifier and the timebase generator inoperative A s...

Page 19: ...x 8 5V all sections of most semiconductors can be tested without damage However checking the breakdown or reverse voltage of high voltage semiconductors is not possible More important is testing compo...

Page 20: ...cuit which is not defective The test patterns on page M 18 show some typical displays for in circuit tests Miscellaneous A posifive going sawtooth voltage of approximately SV coincident with display s...

Page 21: ...citor 33 vF Z diode under 8V Z diode beyond 12V Silicon diode Germanium diode Rectifier Thyristor G A together Junction E C FET In circuit Semiconductors Diode paralleled by 6800 2 Diodes antiparallel...

Page 22: ...ring TRIG selector switch in N position Trigger coupling selected with TRIG switch AC DC LF HF Trig freq range AC and DC to 20 MHz HF above 20 MHz LF below 50 kHz Video signals with line freq TV SEP s...

Page 23: ...G The selection of a cathode ray tube without any tolerances is practically impossible Astigmatism Check Check whether the horizontal and vertical sharpness of the display are equal This is best seen...

Page 24: ...n coefficients on all positions of the input attenuators and readjust them as necessary A compen sated 2 1 series attenuator is also necessary and this must be matched to the input impedance of the os...

Page 25: ...e visible If both vertical inputs are AC coupled to the same signal and both traces are brought to coincide exactly on the screen when working in the alternate dual channel mode then no change in disp...

Page 26: ...l the nearest edge of the calibration signal can be brought in the screen center and checked on the above mentioned criteria With x200 expansion the display brightness nor mally needs increasing with...

Page 27: ...4cm SOmV cm the Y output can be calculated as 2 modulation optional Checking the blanking facility on the 2 modulation connec tor rear panel is possible with a sine or square wave generator The sine w...

Page 28: ...mally the capacitors are discharged 6 sec onds after switching ofi However with a defective instrument an interruption of the load is possible Therefore atter switching off it is recommended to con ne...

Page 29: ...y deviations of the specified data or func tions The HAMEG Service Department can give you advice for troubleshooting and replaceable parts Replacement parts can be ordered by letter or telephone from...

Page 30: ...e safety earth pin via long sold tab to rear chassis Safety Class I Plug with earthing contact I bk II 1 _ 2 tl J 2 gr Pushbutton power switch 4At24A _ 250v see diagram POWER SUPPLY Adjustments As adv...

Page 31: ......

Page 32: ......

Page 33: ......

Page 34: ......

Page 35: ......

Page 36: ......

Page 37: ......

Page 38: ......

Page 39: ......

Page 40: ......

Page 41: ......

Page 42: ......

Page 43: ......

Page 44: ......

Page 45: ......

Reviews: