![Hamamatsu Photonics C9100-02 Instruction Manual Download Page 25](http://html1.mh-extra.com/html/hamamatsu-photonics/c9100-02/c9100-02_instruction-manual_3400153025.webp)
C9100-02 Ver.1.7
(2) Level trigger mode (pulse polarity is negative)
Trigger In cycle > Exposure time + Readout time
Over 100 µs
Within 64 µs
Exposure
Trigger In
Trigger In cycle
Exposure
Exposure time + Readout time
Readout
Readout
Exposure
In level trigger mode, exposure time is set by width of Trigger In.
Trigger In cycle < Exposure time + Readout time (Cyclic trigger mode)
Over 100 µs
Within 64 µs
Exposure
Trigger In
Trigger In cycle
Exposure
Exposure time + Readout time
Readout
Readout
Exposure
Readout
In level trigger mode, exposure time is set by the width of Trigger In.
Readout
23