background image

 

 

Notes

 

7100.006204 PS17 Instruction manual

 

22

 

08.04.19 

 

 

Summary of Contents for PS17

Page 1: ...Instruction manual PS17 differential pressure transmitter...

Page 2: ...ust not be copied either wholly or in part or made available to third parties The instruction manual is part of the product Please read this manual carefully follow our instructions and pay special at...

Page 3: ...7100 006204 PS17 Instruction manual 3 08 04 19 Table of Contents...

Page 4: ...ly be cleaned from the outside with a damp cloth 1 2 Warning symbols The device has been designed and tested to ensure its safety However it may still be dangerous if used inappropriately Precautions...

Page 5: ...e label The piezoresistive pressure transmitter can be used to measure symmetrical or asymmetrical measurement ranges 4 2 Models The device is available in different models With a fixed measurement ra...

Page 6: ...ithout display b Model with display 4 4 Internal control units 1 DIP switch SW 1 and 2 set switchable measurement ranges optional otherwise no function SW 3 5 set the output signal 2 Terminal Connecti...

Page 7: ...in the following order 1 program version 2 final value of the selected measurement range for approx 1 5 seconds 3 measured values 1 Output value displays pressure in Pa and kPa root extracted signal...

Page 8: ...website https www halstrup walcher de en products all PS17 php 5 1 Mounting on a top hat rail and dismantling You can mount the device on a DIN rail see assembly instructions Please note the dimensio...

Page 9: ...rical asymmetrical higher pressure at port lower pressure at port 0 125 Pa e g via a primary element e g a measuring orifice 200 Pa 300 Pa 1 kPa Do not allow hoses to kink or sag Avoid accumulating mo...

Page 10: ...er the digital zero point calibration input is to be used and inform the installation technician Always observe the operating requirements indicated on the type label and in the data sheet particularl...

Page 11: ...ssigned Ground for supply voltage or output signal Ground for supply voltage 3 connection Output signal voltage current Output signal voltage current 4 not assigned Ground for supply voltage or output...

Page 12: ...ing parts if it slips shakes wrong connection This protects the device and persons against contact with the 230 V supply in the event of a fault Note on the circuit board a warning symbol has been inc...

Page 13: ...isplay you will require a display device to read the voltage signal 2 10 V or current signal 4 20 mA If available you can also use the optional display 1 Disconnect the device from the power supply 2...

Page 14: ...nt calibration is complete after approx 5 6 seconds The device is now zeroed Note You can check that the zero point calibration has been performed successfully by measuring the output voltage For mode...

Page 15: ...sion must be configured by the manufacturer output signal always 4 20 mA Only the root extracted signal settings can be adjusted Requirements Open the housing see assembly instructions To make the set...

Page 16: ...t is displayed in and calculated using the following function Root extracted signal Measured value Final value of measurement range The value is displayed to one decimal point The following table expl...

Page 17: ...rted For devices with an integrated display the final value for the selected measurement range will be displayed for approx 1 5 seconds after the restart 6 The measurement range is now set 7 Perform t...

Page 18: ...ebsite https www halstrup walcher de en products all PS17 php Please contact us if you require any further information 9 Scope of delivery The following parts are included in the scope of delivery 1 P...

Page 19: ...tive Send the device to the manufacturer for repair Output signal is constant despite pressure change Output protection diode defective Send the device to the manufacturer for repair Pressure ports re...

Page 20: ...Declaration of Conformity 7100 006204 PS17 Instruction manual 20 08 04 19 12 Declaration of Conformity...

Page 21: ...Notes 7100 006204 PS17 Instruction manual 21 08 04 19 Notes...

Page 22: ...Notes 7100 006204 PS17 Instruction manual 22 08 04 19...

Page 23: ...Notes 7100 006204 PS17 Instruction manual 23 08 04 19...

Page 24: ...Tel 49 7661 39 63 0 info halstrup walcher com www halstrup walcher com halstrup walcher GmbH Stegener Stra e 10 79199 Kirchzarten Germany...

Reviews: