P26 Instruction Manual
29
Troubleshooting
Problem
Cause
Corrective Action
Instrument does not work;
nothing on display
No power
Check the terminal connections
and supply voltage
Only m³/s,m³/h, ft3/h, kg/s,
kg/min and kg/h may be
used as display units.
Air consumption function
activated.
Delete consumption sum and
deactivate air consumption
function.
Scaling menu does not
appear in the menu
Air consumption function
activated.
Delete consumption sum and
deactivate air consumption
function.
Air consumption function
does not appear in the
menu
Incorrect display unit.
Only m³/s, m³/h,
ft3/h, kg/s, kg/min and kg/h
may be used with this
function.
Please select the correct display
unit
Air consumption function
does not appear in the
menu
Air consumption deactivated
Please contact manufacturer
„Na“ is displayed as
current value
Wrong settings for scale
Change scale
Pressure drops
continuously
Leak
Firmly slide tubing completely
onto ports; adjust diameter
No serial communication
No cable connection
Make sure cable is connected
securely