EMA 200 Instruction Manual
11
5 Configuration
By adjusting the configuration settings, users can adapt the instrument to their
specific application. This menu can be activated by depressing the zero-point
key until
'ConF
' appears in the top line of the display. After the operator
releases this key, the instrument launches the first item (setting units of pressure). To
exit the configuration menu, simply press the zero-point key again until '
MEAS
'
appears in the top line of the display. The new settings are now saved and can be
used each time the instrument is switched on.
5.1 Setting units of pressure
The symbol '
[P]
' will appear in the bottom line of the display and one of the pressure
units will either be displayed or indicated as active by means of an arrow.
Users may now set the desired units by pressing the and keys. Briefly
depressing the key will exit this mode and the instrument will continue on to the
next configuration parameter.
5.2 Temperature display
The symbol '
°C
' is shown in the bottom line of the display and the current status of
this switch (on or off) is shown in the top line. Users may change the
status by depressing the and keys, and may finish entering
the settings by pressing the key. The temperature display will automatically
switch off if, during the subsequent step, the flow rate display is activated.
(The instrument can only display temperature or flow rate, but not both at the same
time.)
5.3 Flow-rate display (not available with 200kPa Range)
'SPd' (
SP
ee
d
) will now appear in the bottom line of the display to indicate the function
that controls the flow-rate display. The current status of this switch (on
or off), shown in the top line, may be altered by pressing the and keys.
Briefly depressing the key will exit this mode.
If the flow-rate display has been activated, the menu for setting the
density of the medium to be measured appears; if not, the next step is
to set the scale for the analogue output (section 5.6).