4.2 Air blower and heater assembly.
The air blower will help to keep the window clear of dust, water and light snow. An internal heater can be switched on
(the red button), which is controlled via a thermo-switch, which will enable the heater during temperatures lower than
approximately 5ºC. The fan operation is via the blue button.
The blower unit is secured onto the back of the window mount with screw that came with the blower, and a dowel
locates the assembly. It is powered from the UPS box’s ‘Aux’ output and connected using the supplied cable.
The heater draws approximately 20 Watts.
Locating
dowel
Connector
cover