Maya2000Pro Series Spectrometers 
Installation and Operation Manual 

 

 

 

 
 

For Products: Maya 2000Pro and  
Maya2000Pro-NIR 
Document: 020-00000-000-02-201603b 
 
 

 

 

Summary of Contents for Ocean Optics Maya2000Pro Series

Page 1: ...000 00000 000 02 A Maya2000Pro Series Spectrometers Installation and Operation Manual For Products Maya 2000Pro and Maya2000Pro NIR Document 020 00000 000 02 201603b...

Page 2: ...from Ocean Optics Inc Trademarks All products and services herein are the trademarks service marks registered trademarks or registered service marks of their respective owners Limit of Liability Ever...

Page 3: ...Software 3 Sampling System Overview 3 How Sampling Works 3 Modular Sampling Accessories 4 Interface 4 Shipment Components 4 Other Accessories Available 5 Breakout Box 5 Chapter 2 Installing the Maya20...

Page 4: ...librating the Wavelength of the Spectrometer 16 Saving the New Calibration Coefficients USB Mode 18 Appendix B Specifications 19 Overview 19 How the Maya2000Pro Series Works 19 Maya2000Pro Series Comp...

Page 5: ...options and shipment components Chapter 2 Installing the Maya2000Pro Series Provides installation and configuration instructions Chapter 3 Troubleshooting Contains recommended steps to isolate and co...

Page 6: ...contact Ocean Optics for specific instructions when returning a product Our 3 Year Warranty covers Ocean Optics miniature fiber optic spectrometers light sources and sampling accessories regardless o...

Page 7: ...ence when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction ma...

Page 8: ...About This Manual vi 020 00000 000 02 201603b...

Page 9: ...s Other performance advantages of this detector include great signal to noise characteristics low dark current and good signal processing speed The Maya2000Pro Series onboard module has 10 user progra...

Page 10: ...high sensitivity in the NIR range while the Maya2000Pro offers high sensitivity in the UV range See Maya2000Pro Series Spectrometers Specifications for a side by side comparison Features Hamamatsu S1...

Page 11: ...ues directly from the spectrometer enabling the ability to hot swap spectrometers between computers without entering the spectrometer coefficients manually on each computer About OceanView Software Oc...

Page 12: ...replacing an accessory Interface You can use the Maya s USB connectivity with any computer that meets the following requirements Microsoft Windows Windows 2000 XP 7 32 bit and 64 bit and Windows Vist...

Page 13: ...rs Lithium Ion Battery Pack HR4 BREAKOUT Breakout Box Breakout Box Ocean Optics also offers the Breakout Box HR4 BREAKOUT a passive module that separates the signals from their 22 pin port to an array...

Page 14: ...1 Introduction 6 020 00000 000 02 201603b...

Page 15: ...es Installation This section contains instructions for connecting the Maya2000Pro Series Spectrometer to a computer Procedure Follow the steps below to connect the Maya2000Pro Series to a computer via...

Page 16: ...should be a dynamic trace displayed in the bottom of the graph If you allow light into the spectrometer the graph trace should rise with increasing light intensity This means the software and hardwar...

Page 17: ...ck on the Add Remove Controls tab 2 Select GPIO The GPIO controls appear on the Main Controls tab 3 Check the Out box for Index 4 This makes the GPIO an output 4 Check the Value box for Index 4 to clo...

Page 18: ...ructions asp Change the Slit There s no need to calibrate your spectrometer when changing the slit just install and start measuring There is one exception to this You cannot change from a standard sli...

Page 19: ...ng the slit to the spectrometer 3 Pull the slit out of the spectrometer 4 Put the new INTSMA slit connector into the spectrometer with the key of the connector on the left side 5 Install the 2 screws...

Page 20: ...2 Installing the Maya2000 Series 12 020 00000 000 02 201603b...

Page 21: ...hat you must correct before your Ocean Optics device will operate properly Follow the applicable steps below to remove the incorrectly installed device device driver and installation files Note If the...

Page 22: ...ct the Maya2000Pro Series from your computer 6 Plug in the spectrometer The system is now able to locate and install the correct drivers for the USB device Mac Operating Systems Since there are no dev...

Page 23: ...of time and environmental conditions Ocean Optics recommends periodically recalibrating the Maya2000Pro Series About Wavelength Calibration You are going to be solving the following equation which sho...

Page 24: ...sisTookPak VBA Calibrating the Wavelength of the Spectrometer Procedure Perform the steps below to calibrate the wavelength of the spectrometer 1 Place the spectrometer operating software into Scope m...

Page 25: ...10 0 06 0 13 0 19 5 Use the spreadsheet or calculator to calculate the wavelength calibration coefficients In the spreadsheet program find the functions to perform linear regressions If using Quattro...

Page 26: ...un the Setup exe file to install the software The Welcome screen appears 5 Click the Next button The Destination Location screen appears 6 Accept the default installation location or click the Browse...

Page 27: ...pectrometer operation specifications and system compatibility It also includes accessory connector pinout diagrams and pin specific information How the Maya2000Pro Series Works The optical bench has n...

Page 28: ...ians can change the Filter 4 Collimating Mirror Focuses light entering the optical bench towards the Grating of the spectrometer Light enters the spectrometer passes through the SMA Connector Slit and...

Page 29: ...o Maya2000Pro NIR Detector Hamamatsu S10420 back thinned FFT CCD Hamamatsu S11510 back thinned FFT CCD Thermoelectric cooling No Number of pixels All 2068 x 70 Active 2048 x 64 Spectral range 200 1100...

Page 30: ...ght 0 96 kg 2 12 lbs Temperature Operation Storage 0 C to 50 C 30 C to 70 C Humidity 0 90 noncondensing Power consumption 500 mA 5 VDC Supply Voltage 4 5 5 5 V Power up Time 2s depending on code size...

Page 31: ...trand optical fiber 0 22 NA Integration time 6 ms to 5 seconds Interfaces USB 2 0 System Compatibility You can use the Maya2000Pro Series USB connectivity with any computer that meets the requirements...

Page 32: ...ectrometer pin numbering is as follows USB Port 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 30 Pin Accessory Connector Pinout Diagram 30 Pin Accessory Connector Pi...

Page 33: ...pin can power other peripherals Ensure that peripherals comply with USB specifications no TEC power 13 SPI Data Out Output SPI Master Out Slave In MOSI signal for communication to other SPI periphera...

Page 34: ...oftware programmable digital input output channel number 27 Ground Input Output Ground 28 GPIO 8 Input Output General purpose software programmable digital input output channel number 29 Ground Input...

Page 35: ...ure 16 calibration coefficients saving in USB mode 18 CCD 21 collimating mirror 20 compatibility 23 Components Table 20 configuring in OceanView TE Cooler 8 D detector 21 document audience iii purpose...

Page 36: ...software OceanView 3 specifications 19 detector 21 spectrometer 22 specifications 21 spectroscopic accessories 10 System Requirements 3 T TE Cooler 8 troubleshooting Linux systems 14 Mac systems 14 Tr...