2: Set-up and Operation 

Set-up for Operation to Control Power Through the Front 
Panel 

Use these set-up instructions when you want to control the laser power through the unit’s front panel 
(Models LASER-785-LAB-ADJ and LASER-785-LAB-ADJ-S).  

 

Procedure 

1.

 

Make sure that the master Power key is in the vertical (“Off’) position. 

2.

 

Connect the 120V AC power plug to the power connector on the rear of the unit. 

3.

 

Plug the other end of the power cord into a standard 110/120V electrical outlet. 

4.

 

Connect an SMA 905 connectorized multi-mode fiber patch cord to the SMA bulkhead connector 
on the front panel of the unit. 

 

5.

 

Note that this laser is configured with a remote interlock on the rear panel. The laser has an RJ-11 
(4-pin telephone jack) that is prewired to provide a closed loop connection for the interlock. 
Removal of the RJ-11 connector or a break in the closed loop circuit provided will result in a 
laser interrupt and the system will be reset to Standby mode. You can reconfigure the RJ-11 
connector to provide a closed loop that interactively couples with their system or facility by 
replacing the short length of wire with longer closed loop connections, if desired. 

6.

 

Ensure that the 

Emergency Shut-off

 switch (EMO) is in the “ready” position by turning the red 

button clockwise until it pops out slightly. 

7.

 

You are now ready to turn on the laser module. Go to 

Operation Controlling Power Through the 

Front Panel

LASER-785-LAB-02-0409 

7

 

Summary of Contents for Ocean Optics LASER-785-IP-LAB

Page 1: ...ouglas Ave Dunedin FL USA 34698 Phone 727 733 2447 Fax 727 733 3962 8 a m 8 p m Mon Thu 8 a m 6 p m Fri EST E mail Info OceanOptics com General sales inquiries Orders OceanOptics com Questions about o...

Page 2: ...se without written permission from Ocean Optics Inc This manual is sold as part of an order and subject to the condition that it shall not by way of trade or otherwise be lent re sold hired out or oth...

Page 3: ...Keep all unauthorized personnel out of the area where the laser is operated to avoid potentially hazardous exposure to eyes and skin 3 Whenever the laser is running and the beam is not in use it is g...

Page 4: ...ty are the following Center for Device and Radiological Health Office of Compliance 2098 Gaither Rd Rockville MD 20850 Tel 301 594 4654 Fax 301 594 4672 Laser Institute of America 12424 Research Parkw...

Page 5: ...apter 2 Set up and Operation 5 Overview 5 Set up 5 Set up for Operation in Internal Power Preset Mode 6 Set up for Operation to Control Power Through the Front Panel 7 Set up for Operation in External...

Page 6: ...Table of Contents ii LASER 785 LAB 02 0409...

Page 7: ...optics com Select Technical Operating Instructions then choose the appropriate document from the available drop down lists Or use the Search by Model Number field at the bottom of the web page You can...

Page 8: ...About This Manual iv LASER 785 LAB 02 0409...

Page 9: ...laser module that contains an integral Thermo Electric Cooler TEC and thermistor for monitoring the internal temperature of the laser platform The laser has narrow spectral lines is compact and featu...

Page 10: ...spectrum stabilized multi mode laser with 350mW of SMA coupled adjustable output power Also features an LED display and an output power control dial In addition the LASER 785 IP OEM laser is availabl...

Page 11: ...ls LASER 785 LAB ADJ and LASER 785 LAB ADJ S only Package Contents One 1 Ocean Optics Multimode Stabilized Laser One 1 Thermistor One 1 power cord Two 2 Master Power keys One 1 remote plug Additional...

Page 12: ...1 Introduction 4 LASER 785 LAB 02 0409...

Page 13: ...mpting to install and operate your laser Set up The laser set up procedure differs depending on whether you are going to control the output power automatically via the Internal Preset mode Model LASER...

Page 14: ...e interlock on the rear panel The laser has an RJ 11 4 pin telephone jack that is prewired to provide a closed loop connection for the interlock Removal of the RJ 11 connector or a break in the closed...

Page 15: ...t this laser is configured with a remote interlock on the rear panel The laser has an RJ 11 4 pin telephone jack that is prewired to provide a closed loop connection for the interlock Removal of the R...

Page 16: ...the Laser On switch and assuring that the Laser On LED on the front panel is not lit 2 Set the Mode Selection switch on the rear panel to External the amber Operating Mode LED will light 3 Connect a B...

Page 17: ...ation Controlling Power Through the Front Panel Use this procedure to control the laser power through the unit s front panel Models LASER 785 LAB ADJ and LASER 785 LAB ADJ S Procedure 1 Turn the maste...

Page 18: ...nt on the DC bias voltage that has been applied to the Control port on the rear panel Using the Remote Interlock Feature The Remote Interlock is a safety feature that automatically shuts down the lase...

Page 19: ...On toggle switch on the front panel to turn the laser off Make sure that the Laser On LED goes off 2 Turn the master Power key on the rear panel 90 degrees to the vertical off position The green Syst...

Page 20: ...2 Set up and Operation 12 LASER 785 LAB 02 0409...

Page 21: ...tion 1 M manual operation set up 6 7 9 models 2 O operation 8 Controlling Power Through the Front Panel 9 Extarnal Power Control mode 10 Internal Power Preset mode 8 P package contents 3 product relat...

Page 22: ...Index 14 LASER 785 LAB 02 0409...

Reviews: