HG-1-CAL Installation and Operation Instructions
008-00000-STD-01-20100412
5
Strong Hg Emission Lines by Wavelength (nm)
There are more emission lines shown below than printed on the label on the HG-1 housing. The label
is intended as a quick, convenient reference and does not list every Ar and Hg emission line that
exists.
253.652
404.656
296.728
407.783*
302.150
435.833
313.155
546.074**
334.148
576.960
365.015
579.066
* This spectral line is not evident with spectrometers configured with 300 or 600 lines/mm gratings.
** Spectrometers with 1200, 1800, 2400 or 3600 lines/mm gratings have spectral lines evident at
576.96 nm and 579.07 nm.
Strong Ar Emission Lines by Wavelength (nm)
696.543
800.616*
706.722
811.531
714.704
826.452
727.294
842.465
738.398
852.144
750.387
866.794
763.511
912.297
772.376
922.450
794.818
* This spectral line is evident only with spectrometers configured with 1800, 2400, or 3600 lines/mm
gratings.