Chapter 3   

Operating Instructions 

Operating the Lamp

 

The following sections provide instructions on operating the deuterium and halogen lamps in the 
DH-3PLUS Light Source. The DH-3PLUS unit must be in a horizontal position for it to work.  

Starting Main Power 

To apply main power to the unit,  

1.  Ensure that the Main Power Switch is OFF. Then, connect the power cord to the 

Terminal Input on the rear panel of the unit.  

2.  Turn the Main Power Switch ON. The Power LED lights, indicating that the unit is 

receiving power. The Status LED blinks momentarily after the main power is switched 
on. The deuterium and halogen bulbs remain off after power-on; they must be turned on 
separately. 

 

Manual Operation of Lamp  

Starting the Lamp 

Press the DEUTERIUM or HALOGEN On/Off switch down to ignite the desired lamp. 
The LED blinks green until the lamp illuminates (approximately 20 seconds for the 
deuterium lamp, 1-2 seconds for the halogen lamp.) 
 

 

Deuterium Status LED

 

Halogen Status LED

 

Halogen On/Off Switch

 

Deuterium On/Off 

Switch

 

 

Summary of Contents for Ocean Optics DH-3PLUS-CAL

Page 1: ...Deuterium Halogen Calibration Light Source Installation and Operation Manual For Products DH 3PLUS CAL DH 3PLUS BAL CAL Document 355 00000 003 01 Version 1 0...

Page 2: ...permission from Ocean Optics Inc Trademarks All products and services herein are the trademarks service marks registered trademarks or registered service marks of their respective owners Limit of Liab...

Page 3: ...afety Notices vi Warranty vi Certifications and Compliance vii Chapter 1 1 Setup 1 Overview 1 Models Available 2 Unpacking the DH 3PLUS Series Light Source 2 Contents 2 Connecting a Spectrometer to th...

Page 4: ...p Off 12 Warming Up the Lamp 12 Manual Operation of the Shutter 12 Remote Operation of the LAMP and Shutter 13 Chapter 4 14 Troubleshooting 14 Bulb Replacement 15 Appendix A 16 Calibration Basics 16 O...

Page 5: ...Specifications Contains operating environment specifications as well as other physical details of the product Chapter 3 Operating Instructions Provides instructions for operating the DH 3PLUS lamp and...

Page 6: ...le parts inside 3 Only allow qualified personnel to operate this unit 4 Do not use the unit if it is damaged in any way Contact your dealer for repair or replacement information 5 Always screw in the...

Page 7: ...be disposed of by arranging to return to a designated collection point for the recycling of waste electrical and electronic equipment Separating and recycling this waste equipment at the time of dispo...

Page 8: ...nst harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance wit...

Page 9: ...rd traceable to the National Institute of Standards and Technology NIST it is specifically designed for calibrating the absolute spectral response of your system Calibrated sources are guaranteed for...

Page 10: ...00 250 2400 Yes Cosine Corrected Bare Fiber The following sections provide instructions on unpacking and setting up your DH 3PLUS Series Light Source Unpacking the DH 3PLUS Series Light Source Procedu...

Page 11: ...o a spectrometer 1 Loosen the set screw on the DH 3PLUS s SMA connector 2 Screw the CC 3 cosine corrector onto the end of the fiber The connection should be tight Insert the CC 3 all the way into the...

Page 12: ...record a new dark spectrum before each measurement This minimizes the influence of a shifted dark spectrum Use a log to monitor the operation time of the light source Recalibration of the light sourc...

Page 13: ...L DH 3PLUS CAL Bare Fiber DH 3PLUS BAL CAL DH 3PLUS BAL CAL DH 3PLUS CAL DH 3PLUS CAL 230 1100 standard 230 2400 extended 210 1100 standard 210 2400 extended 250 1100 standard 250 2400 extended 250 11...

Page 14: ...ut condensation at 40 C Ventilation Unit should be situated so that its location or position does not interfere with proper ventilation Temperature 5 C 35 C Unit should be situated away from any devic...

Page 15: ...Covered to protect users from unintentionally looking directly at the beam of light Connect the fiber cable to the DH 2000 BEFORE turning the lamp on to avoid unnecessary exposure to UV radiation Alw...

Page 16: ...le to the lamp when the Main Power Switch is in the OFF position THE INSTRUMENT MUST BE CONNECTED TO A GROUNDED EARTHED OUTLET TTL Input D SUB 15 pin connector for automatic shutter control and remote...

Page 17: ...val below and the date of the calibration as well as the serial number of the light source The calibration acronyms are defined in the following table Calibration Acronym Description CC Cosign Correct...

Page 18: ...briefly at power up 6 NA 7 NA 8 D_BAD In the event the Deuterium lamp is bad this will be active low otherwise it will be hi This equates to the Red Deuterium LED on the front 9 D_GOOD This is active...

Page 19: ...00nm Littlefuse 0218002 Fuse DH CAL Calibration service for DH light sources DH 3P CAL 210 1100nm DH 3P BAL CAL 230 1100nm DH CAL BF Bare fiber calibration service for DH sources DH 3P CAL 250 1100nm...

Page 20: ...panel of the unit 2 Turn the Main Power Switch ON The Power LED lights indicating that the unit is receiving power The Status LED blinks momentarily after the main power is switched on The deuterium...

Page 21: ...t beam as this can cause eye damage Turning the Lamp Off Turn the Deuterium or Halogen lamp off by tapping the appropriate ON OFF switch Wait 60 seconds before powering off the unit Warming Up the Lam...

Page 22: ...t on your spectrometer USE ONLY THE CABLE PROVIDED OR A CUSTOM CABLE DESIGNED TO INTERFACE WITH THE PINOUTS AS DESCRIBED IN THE PINOUT INFORMATION SECTION DO NOT USE A VGA VIDEO CABLE Set the front pa...

Page 23: ...ff switch lights up red indicating an error Deuterium lamp too hot Allow the Deuterium lamp to cool down 20 minutes Press On Off switch again to reset the Deuterium Tungsten Halogen lamp then press ag...

Page 24: ...en lamp LED does not light after switching on the Halogen lamp Internal power supply is defective Disconnect the unit from the main power source and contact your dealer for repair or replacement Bulb...

Page 25: ...irradiance has the physical unit W nm cm 2 The interpretation is as follows This is the wavelength resolved electromagnetic radiation which is emitted through an area cm 2 When you attach a cosine cor...

Page 26: ...ce for the calibration in the NIR for length larger than 1100 nm 3 d Uncertainty in mounting distance of the detector probe e g Cosine Corrector 1 8 per 0 1 mm integrating sphere minimal bare fiber sa...

Page 27: ...9 6 5 6 0 4 1 3 9 3 8 4 3 5 6 10 6 11 7 k 2 15 4 13 8 13 0 12 0 8 2 7 7 7 7 8 5 11 2 21 2 23 5 For DH 3P BAL CAL Bare Fiber Calibration Uncertainty Budgets in at Wavelength Wavelength nm UV Calibrati...

Page 28: ...ries Use Log Use Log No Page No ID No Lamp S N Date Calibrated Reference Standard Working Standard Note Do not exceed 50 hours of total ON time without recalibration Date Used Start Time Stop Time Pri...

Page 29: ...B DH 3PLUS Use Log 000 10000 191 02 201601 21 Date Used Start Time Stop Time Printed Initials Signature Running Total for on Time Comments...

Page 30: ...L lamp starting 11 turning off 12 warming up 12 lamps 11 log 19 M models 2 O operating environment 5 P package contents 2 parts list 10 pinouts 9 product related documentation v S setup 1 shutter 12 a...

Page 31: ...B DH 3PLUS Use Log 000 10000 191 02 201601 23...

Reviews: