Get Started:

This box includes:

• CODA Coriolis mass flow controller

• NIST-traceable calibration certificate 

• Optional accessories

 

(cables, power adapters, filters)

Install

Install the device into 

your process. Ensure the 

flow path is the same 

direction as the flow 

arrow shown on your 

unit. The device may be 

oriented in any direction.

Signal Output

See reverse side for pinout details

Analog: 

Signal outputs are 

available via 0–5 Vdc (0–10 

Vdc or 4–20mA options)

Digital:

 USB-C, RS-232, 

RS-485 and MODBUS RTU 

over RS-232 or RS-485.

DOC-MANUAL-KC-QUICK-US 

Rev 0 • 2020-11-05

Download the full manual: 

alicat.com/coda

Troubleshooting • Questions:

 

+1 520-290-6060 • [email protected] • Live web chat at alicat.com

Power

Supply power via the 

DB-15 connector. The 

controller requires 

12–30 Vdc. A green 

LED light indicates 

power input.

Analog: 

Apply a 

ground to Pin 12 for 

at least 1 second. 

(See reverse side for 

full pinout details.)

Digital: 

For serial 

tare options including 

MODBUS and 

LabVIEW, consult 

the full manual.

Button:

 Press and 

hold the tare button 

on top of the device 

for 10 seconds.

2

3

Zero (Tare)

After completing step 5, tare the device with one of three methods:

4

Although the device is 

immediately functional, 

allow for 15 minutes of 

warm-up for best accuracy.

5

Flow the fluid to be measured 

through the device, then estab-

lish a no-flow condition while fluid 

is present throughout the unit.

6

1

12

Ground

to tare

DB-15 Male