Maintenance
Maintenance
Cleaning
This instrument requires minimal maintenance. If necessary, the outside of the instrument
can be cleaned with a soft dry cloth. Avoid excess moisture or solvents.
The single most important thing that affects instrument life and accuracy is the quality of
the gas being measured. The instruments are designed to measure clean, dry, non-corrosive
gases.
∆
Caution:
If you suspect that debris or other foreign material has entered your
instrument,do not take apart the flow body to clean it, as this will negate its NIST-
traceable calibration. Please contact support for cleaning (
).
Recalibration
BASIS 2 instruments are calibrated to NIST-traceable standards at the time of manufacture.
As an OEM instrument, there is not a factory-recommended periodic recalibration cycle.
Recalibration can be requested at the user’s discretion/requirement. A recalibration can be
requested with your serial number at
alicat.com/service
.
For repair, recalibration, or recycling of this product contact us (
) with your
instrument’s serial number and your contact information.