13.

 

Carefully rotate the top half of the unit to re-assemble the two halves and give you a clear view 
of the LCD display. 
 

14.

 

Connect the test leads of the Yokogawa 2786 Decade Resistance Box or equivalent device to 
the HAKKO FG-101 where the one test lead is connected to the test plate of the HAKKO FG-101 
and the other test lead is connected to the grounding terminal of the HAKKO FG-101. 
 

15.

 

Verify the calibration by setting the output value on the Yokogawa 2786 Decade Resistance Box 
or equivalent device to the values in the table below and confirm that the display shows the 
standard values corresponding to each output setting.  
 

Output Value 

(

Ω

Standard Value 

(

Ω

0.0 to 0.1

 

1.9 to 2.1 

40 39.5 

to 

40.5 

 

16.

 

If there is any deviation outside the standard values, perform the calibration again. 
 

17.

 

When finished, turn off the HAKKO FG-101.  Remove the test leads of the Yokogawa 2786 
Decade Resistance Box or equivalent device from the HAKKO FG-101.  Re-secure the two halves 
of the unit by installing the 5 Phillips screws along the front and rear of the casing.  
 