background image

Operation

Temperature Recorder (optional)

The blood bank refrigerator is equipped with a 6-inch temperature recorder for the chamber
temperature. When the unit is powered up, the recorder starts working.

For proper operation of the temperature recorder, please follow the procedures below.

a)Open the recorder door to access the recorder.
b)Plug in the 9 volt battery located on the upper right corner of the recorder. This battery is a
standby power.
c)Install a new chart by following the diagram below.

1.Installation and Operation

(Figure 6) Chart Adjustment Buttons

9 volt battery

Chart Adjustment Buttons

Imprinting Stylus

Date Reference Mark

Center Nut and Chain

Chart Paper

(Figure 5)

2.Power Supply

Replacement of Recorder Battery

The recorder normally uses AC power supply as the refrigerator operates.

If the AC power supply

malfunctions, the LED indicator flashes to alert a power supply failure.

The recorder continues to

operate on its standby battery to track the refrigerator temperature. The typical battery can sustain the
chart operation for about 30 hours. Please be aware that a weak battery should be replaced to avoid
corrosion in the connector buttons. The battery power can be saved by disconnecting the connector
button when the unit is not in service. During normal usage, make sure that the battery is plugged in to
power the recorder in case of an AC power outage. The LED indicator glows continuously until the main
power supply is on and the standby battery is replaced. When the battery is low, the LED indicator
flashes to indicate that the battery needs to be replaced.

When the green LED on the recorder flashes, there are two ways to inspect the unit.

Operation

Control Panel

1)Temperature Indicator

This refrigerator is factory preset to run at 5

. The high temperature alarm is set at 10

, the low

temperature alarm is set at 2

.

1

2

3

4

5

Operation

Push Button

Display

Definition

of display

Plug in the unit, power is on

Push temperature display
button

Push temperature display
button

Push temperature display
button

Repeat Step 2

Average Temperature in

Sensor Bottle

Upper sensor bottle

temperature

Lower sensor bottle

temperature

Average Temperature in

Sensor Bottle

Average
Temperature

Upper sensor
bottle
temperature
Lower Sensor
bottle
temperature

Average
Temperature

Note:

The digital display indicates either the upper bottle temperature or the lower bottle temperature

respectively.

The displayed temperature does not always stay at 5

.

The temperature sensor

represents the average temperature inside the refrigerator.

Defrost Cycle

This blood bank refrigerator has two automatic defrost cycles.

To maintain a constant temperature inside the blood bank, the refrigeration system cycles on and off

according to the temperature demand. When the refrigeration compressor cycles off, a small electric

heater is energized to thaw the frost off the evaporator surface. The function does not have an obvious

impact to the chamber temperature.

When the environmental humidity is too high or the products loaded into the refrigerator emit much

moisture, cycle defrost will not be sufficient to remove all frost and ice accumulated on the evaporator

surface. The equipment will start the forced defrost cycle. Once the frost is thawed, the refrigerator will

resume normal operation. During the defrost process, the temperature of a 200-ml blood bag would

probably rise by 1~ 2

.

1)Cycle defrost

2)Forced defrost cycle

Table 1, Temperature Indication

Summary of Contents for HYC-610

Page 1: ...Pharmaceutical Refrigerator Operation Manual HYC 610 ...

Page 2: ...on 6 Operation 7 11 Proper Usage 12 Maintenance 13 Specification 14 Service Support and Description of Labels 15 Table of Contents Thoroughly read this manual before using the refrigerator Please keep this manual for convenient reference The actual color and appearance of the refrigerator may vary slightly from catalog ...

Page 3: ...prohibited This symbol designates procedures to be followed Please keep this manual for convenient reference Warning This action may cause fatal injury or severe damage Environmental Conditions This unit is designed for applications in the following environmental conditions It is safe to operate the unit at these or more favorable conditions 1 Indoor usage 2 Maximum relative humidity of 80 percent...

Page 4: ...o dismantle repair or modify the refrigerator Such action may cause fire or injury of personnel Should there be malfunctions of the unit unplug the power cord from the power supply Continuous operation of the unit abnormally may cause fire or electrical shock When used for storage of poisonous products or products that are radiant and products that can be harmful to public health the unit needs to...

Page 5: ...tor alarm buzzer remote alarm contact Flashing Alarm Indicator buzzer alarm without time delay remote alarm contact Alarm for door opening after 10 minutes delay Non Volatile memory storage 15 W X 1 Including one box of recorder chart paper and 4 pieces of 9 V battery optional One set of key one certificate of warranty one copy of operation manual one plastic bag Operating Parameters Model Sensor ...

Page 6: ...g the light bulb starter Purchasing chart paper for temperature chart recorder Purchasing backup battery for temperature recorder Changing fuse breaker 1 The refrigerator should be cleaned once a month Regular cleaning can keep the unit looking new 2 Use a dry soft cloth to clean the dust off the interior and exterior If necessary use a soft cloth with a solution of water and mild detergent to wip...

Page 7: ...on of Refrigerator Your refrigerator should be placed at a location that satisfies the following conditions By doing so the optimum operating results can be achieved 1 Firm level floor surface 2 A place that is far away from any heat source 3 Do not install a refrigerator in direct sunlight 4 Dry area 5 Clean area 6 Good ventilation 7 Ground protection Installing the unit on a firm and flat floor ...

Page 8: ...of the recorder moves to the left side of the chart c When the stylus is completely off the chart unscrew the center nut for the paper and remove the old chart paper off the recorder Place a new chart paper on the recorder Carefully match the day and time line on the chart with the reference mark on the recorder panel there is a small groove on the front of the recorder as shown in the figure d Pl...

Page 9: ...ashes there are two ways to inspect the unit Operation Control Panel 1 Temperature Indicator This refrigerator is factory preset to run at 5 The high temperature alarm is set at 10 the low temperature alarm is set at 2 1 2 3 4 5 Operation Push Button Display Definition of display Plug in the unit power is on Push temperature display button Push temperature display button Push temperature display b...

Page 10: ...ator or a refrigerator that has been unplugged for a long time you should make sure the battery is fully charged This charging process can take up to two days When the battery discharges the remote alarm contact continues to function Procedure to Test the Alarm System Push the Alarm Test button once The alarm buzzer sounds three consecutive times in 1 Hz frequency At the same time the alarm light ...

Reviews: