10-8
Shroud and Base Assemblies
P/N 108755-US, Manual revision: AA
S
HROUD
A
SSEMBLY
I
NSTALLATION
1.
To assemble the shroud assembly to the base, rotate the link assembly un-
til the Link D is in its highest position.
2.
Lay the Shaker in a cradle with the bottom of the Shaker facing up.
3.
Loosen the screws on the rear boss inside the shroud and slide it onto the
link until the shaft in the Link D slides into the grooves in the rear boss of
the shroud and behind the cone screw/washers. Using a long screwdriver,
tighten the rear boss screws.
4.
Apply a small amount of grease in the center of the female knife.
5.
Position the female knife over the knife edge.
6.
Place screws into the holes on each end of the female knife.
7.
Bring the shroud up to the female knife, aligning screws with boss.
8.
Through the bottom of the base, hold one screw in place with a long phil-
lips screwdriver and start the screw.
9.
After beginning one screw, start on the other one and then tighten both.
Summary of Contents for SEBRA 1040
Page 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Page 3: ......
Page 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Page 8: ...vi Preface P N 108755 US Manual revision AA...
Page 12: ...x Table of Contents P N 108755 US Manual revision AA...
Page 14: ...xii List of Figures P N 108755 US Manual revision AA...
Page 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Page 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Page 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Page 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Page 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Page 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Page 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Page 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Page 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Page 105: ......