126
Chapter 7, Protocol Settings
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
Modifiable Settings
Default Settings
displays the Haemonetics default settings for the
Cell Saver Elite device:
Table 11, Default Settings
Cell Salvage
Parameter
Values
Default
Fat Reduction
On/Off
Off
Auto-Fill
On/Off
On
Auto-Wash
On/Off/Ask/
Skip
On
Partial Bowl Wash
Ask/Single/
Double
Ask
Pump Regulation
On/Off
On
a
Empty Destination
Blue/Red
Blue
Volume Accounting
b
Reservoir/
Circuit
Reservoir
Final Cycle
On/Off
Off
Smart Empty
On/Off
On
Parameter
225 mL Bowl
125 mL Bowl
70 mL Bowl
Min
Max
Step
Fill Start Volume (mL)
800
800
400
200
3000
100
Fill Resume Volume (mL)
400
400
200
200
3000
100
Fill Pump Rate
c
(mL/min)
500
225
100
25
1000
25
Minimum Wash Volume
(mL)
1000
750
300
500
d
5000
250
e
Wash Pump Rate
c
(mL/
min)
450
200
100
25
1000
25
Empty Pump Rate
f
(mL/
min)
500/400/100
300/150/100
100
25
1000
25
Emergency Fill Pump
Rate (mL/min)
800
400
N/A
25
1000
25
Emergency Wash Pump
Rate (mL/min)
800
400
N/A
25
1000
25
Emergency Empty Pump
Rate
f
(mL/min)
500/400/250
300/250/100
N/A
25
1000
25