124
Chapter 7, Protocol Settings
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
4. Touch
(
Accept
). The parameter list for the new settings group
appears.
5. Set the parameters for the new settings group, then touch
Done
. For
more information, see “
.”
Editing a
Settings Group
The parameter list displays the individual settings in a settings group. Follow
the steps below to edit the settings in a settings group.
1. Select a settings group to edit.
2. Touch
View
. The parameter list appears.
3. Touch a tab to select a bowl size.
4. From the list, select a setting to modify.
5. Touch
(
Up
) or
(
Down
)
to change the value for that setting.
Parameters that have been changed from the default setting will have a
dot next to the value.
6. To reset a setting to its default value, select the setting and touch
Default
.
7. Touch
Done
to save the changes and return to the
Settings
screen.
Figure 65, Example of the parameter list
Note: If you navigate to a different screen without touching
Done
, your changes
will still be saved. All changes will take place immediately, except target volume