Event Messages
The following are possible event messages related to the connectivity feature
that may occur on the Cell Saver Elite/Elite+ device. Refer to the user manual
for a comprehensive list of event messages.
Table 2, Event Messages
ID #
Event Message Text
502
Software Update Available
Explanation
:
A software update has been downloaded and is available for installation. Operator
confirmation is required before the software can be installed.
Note:
The device cannot be used while the software installation is in progress, which will take
several minutes.
Corrective Action:
Touch
Install Update
to install the software now.
or
Touch
Skip Update
to skip the software installation. This prompt will recur on the next power
up.
503
Settings Update Available
Explanation:
System and protocol settings have been downloaded and are available for application.
Operator confirmation is required before these settings will be applied.
Corrective Action:
Touch
Apply Update
to apply the settings now.
or
Touch
Skip Update
to skip the application of the settings. This prompt will recur on the next
power up.
504
Software Download in Progress
Explanation:
New device software is downloading. The device can be used normally.
505
Settings Applied
Explanation:
New system and protocol settings have been applied.
506
Network Notification
Explanation:
[Customizable message]