Equipment Description
19
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Disposable Sets
Elements of the
Glycerolization
Disposable Set
Figure 2, LN225 Glycerolization disposable set
The glycerolization disposable set contains the following elements:
Glycerol solution line:
The glycerol solution line contains a spike which
is used to connect the glycerol solution bottle to the disposable set and a
ratchet clamp.
Pump tubing:
The blood pump tubing sections contain
two blue
pump
stops which will secure the pump tubing during pump loading and
operation.
Bacteria barrier filter:
A 0.2 µm is integrated on the set, downstream of
the solution pump. This filter acts as a barrier to prevent potential
contaminants from passing into the system after spiking or connecting any
of the solution bags.
Note: The bacteria barrier filter allows the attachment of the glycerol using a
spike, while still maintaining a functionally closed system and allows the
extension of the expiration date of the product. The filter does not restrict the
flow of glycerol.
1.
4.
2.
7.
5.
3.
6.
1. Glycerol solution spike
2. Ratchet clamp
3. Blood pump tubing
4. Bacteria barrier filter
5. DPM filter
6. Slide clamp
7. Heat sealed tubing