134
Disassembly
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
Top Deck Assembly
Top Deck
1. Remove the front and rear panels.
2. Remove the centrifuge assembly. See
.
3. Disconnect the cable connector from the backplane PCB and thread the
cable through the rear panel.
4. Disconnect the rest of the connectors from the backplane PCB.
5. Disconnect the connectors from the top deck distribution board.
6. Disconnect the pneumatic connector from the compressor.
7. Remove Phillips head screw securing the top deck distribution PCB to
the power supply assembly bracket.
8. Disconnect the ground wire at the top deck ground bus that connects to
the side panel ground bus.
9. Remove the (4) 3/8” nuts that secure the top deck to the cabinet.
10.Firmly grasp the top deck and pull away from the cabinet.
Air Detectors,
Deck
Remove the front and rear panels.
Solution Air Detector
1. Disconnect the cable connector from the top deck distribution PCB.
2. Remove the right handle by removing the (2) Phillips head screws
securing the inner and outer panels.
3. Remove the (2) Phillips head screws securing the air detector to the top
deck.
Blood Air Detector
1. Remove the centrifuge assembly.
2. Disconnect the cable connector from the top deck distribution PCB.
3. Remove the (2) Phillips head screws securing the air detector to the top
deck.
Top Deck Air Detector Reassembly
1. Apply a light film of silicone lubricant to the air detector gasket before
attempting to install the air detector.
2. If both top deck air detectors are removed, be sure to return them to the
correct locations. The plugs are keyed, so use the descriptions on the top
deck distribution PCB for identification.