Haehne DMA2 Adjustment Instructions Manual Download Page 4

Operation

Preparation

  - Apply power to the amplifier with connected sensors.  Wait 15 minutes to warm up.
  - Ensure that the completely mounted sensors are loaded with the normal pre-load acting in the regular  

    operating mode without any additional extraneous loads.  In the case of Web tension sensors this is the  

    roll weight without any web (foil, paper, ...)

Adjustment mode

  First press and hold the SET key to switch from the operating mode into the adjustment mode, afterwards  

  press the menu key 

Ü

 briefly and release the SET key.

  The display indicates the 1. menu step which can now be changed.
  For the second, third and any additional menu step do not press the SET key instead use the  

  menu key 

Ü

 to select the desired menu point, for instance, to make changes.

Storing of changes

  Press the SET key until the LED goes OFF.  This activates the adjustment procedure and the LED lights  
  up until the procedure is completed. The amplifier switches now automatically into the operating mode.

Leaving the adjustment mode without storage

  When exceeding the waiting time of 20 seconds the amplifier switches without changes automatically into  

  the operating mode.  An alternative is pressing the menu key beyond the 8. menu step.

Display intensity

 

(7. Menu step) 

 

select the menu step "display intensity"  

 

in the adjustment mode.

 

- select the display intensity of the LEDs

 

- all four LEDs blink during the adjustment

 

- store adjustment

Pre-Adjustments / Operating Modes

page 4

Filter type of the current output

 

(1. Menu step) 

 

select the menu point "filter type"  

 

in the adjustment mode

 

- select the operating mode 

 

- LED blinks during the adjustment

 

- store adjustment

 

- if "FIL" was selected then the LED  

 

  remains ON in the operating mode

increasing

decreasing

HAEHNE

  Elektronische Messgeräte GmbH · Heinrich-Hertz-Str. 29 · D-40699 Erkrath Germany · Telefon 0211/9 25 91-0 · Fax 0211/9 25 91-20

http://www.haehne.de 

Email: [email protected]

HAEHNE

08.08 DE1DM2EA.indd 

Summary of Contents for DMA2

Page 1: ...in 2 Amplification adjustment with calibration weights 2 Connection Wiring 3 Operation Preparation 4 Adjustment mode 4 Pre Adjustments Operationg Modes Display intensity 4 Filter type of the current output 4 Output signal range 5 Display value 5 Adjustement Of The Parameters Automatic zero point adjustment 6 Manual zero point adjustment 6 Entering amplification as gain value 6 Amplification adjust...

Page 2: ...inal rating 1 5 m V V bridge supply voltage 10 V the output signal is therefore 15 mV 0 015 V at nominal force Amplifier output 10 V at nominal web tension 0 V at zero web tension If the sensor is operating below its nominal force then the amplification must be increased Example The maximum sensor load is only 75 of nominal force The output signal is therefore only this leads to Amplification adju...

Page 3: ...y from the shield with PE Additional recommendation Connect the 0 V of the 24 V circuitry at a defined place prefably in close proximity of the PE If this reference point is to be operated in the floating point mode then the voltage between 0 V and PE should not exceed 50 Vpp The four terminal blocks with five terminals each can be separatly plugged in simplifying trouble shooting in case of fault...

Page 4: ...re and the LED lights up until the procedure is completed The amplifier switches now automatically into the operating mode Leaving the adjustment mode without storage When exceeding the waiting time of 20 seconds the amplifier switches without changes automatically into the operating mode An alternative is pressing the menu key beyond the 8 menu step Display intensity 7 Menu step select the menu s...

Page 5: ...ds to 10 V output voltage increasing The various decimal ranges are stepped through sequentially The step velocity increases with the key holding time This adjustment deletes stored peak values Output signal range 8 Menu step select the menu point output signal range in the adjustment mode select operating mode store adjustment Operating Mode Signal value Voltage output Current output Option C Opt...

Page 6: ...sing adjustment range 10 0 100 Entering amplification as gain value 4 Menu step calculate amplification siehe section Adjustment Modes on page 2 if no sensor is connected then terminals 18 19 or 13 14 have to be connected with each other Select the menu point Gain in the adjustment mode select the desired amplification store the adjustment increasing decreasing adjustment range 400 2800 page 6 See...

Page 7: ...nd is displayed in menu step 4 calculated amplification as gain value null The calibration weight is too small Model the web geometry with a cable or rope and apply calibration weight The weight should approxi mate the actual web tension in the operating mode The DMA2 offers the facility to increase the weight with a percentage calculation As an example 70 of the required 100 kg are applied to the...

Page 8: ...the SET key max value min value Fault display during operation Exceeding the maximum output signal voltage of approximately 13 3 V e g resulting from a broken sensor cable HAEHNE Elektronische Messgeräte GmbH Heinrich Hertz Str 29 D 40699 Erkrath Germany Telefon 0211 9 2591 0 Fax 0211 9 2591 20 http www haehne de Email info haehne de HAEHNE page 8 08 08 DE1DM2EA indd ...

Reviews: