6
Safety Instructions
chuck (test this with a dipstick and clean if
necessary),
•
the inserted tool sits properly in the chuck
and will not cant/jam while clamping it,
•
the tool is approved for using it at the maxi-
mum speed (see manufacturer’s instruc-
tions) or only use the speed required by the
instrument,
•
the tool is absolutely dry.
For safety reasons do not change the tool while
the motor is running.
Switch off the unit immediately if
•
the bur comes off during the treatment itself.
•
the motor blocks during the treatment.
While doing any foot care, only operate the unit
with the suction turned on.
Only use the suction function, if you have in-
stalled the suction filter. Check this filter every
10 operating hours. If it is full, please exchange
it. A filter should be exchanged at least after 50
operating hours.
Never use the suction to vacuum any liquids (or
air/water fog which contains particles of liq-
uids).