background image

 

14 

Application 

Insert the bur  

Insert  the  bur  until  it  stops.  The 
hand  piece  is  equipped  with  a 
chuck which can hold all burs with a 
diameter  of  2.35  mm  (confirm  to 
ISO 104).  

 

 

Only  change  the  tool  if  the 
hand piece is turned off. 

 

Adjusting the speed 

 

Put the hand piece back in the hand 
piece holder or keep it in your hand. 
Switch  on  the  unit  with  switch  (8). 
Now  you  can  adjust  the  requested 
speed  with  the  switch  (3)  and  (4). 
Choose  the  direction  with  switch 
(6). 

 
 

Memory buttons m1, m2 

Frequently  used  speed  levels  can 
be  saved.  Therefore,  two  memory 
buttons  are  available.  At  first, 
choose  your  speed  level  and  then 
save  it  by  pushing  button  m1  for  4 
seconds. To recall the above speed 
level simply press button m1 again. 
Further  speeds  can  be  saved  with 
m2. 
 

 

Spray 

Turn  on  the  spray  function  while 
pushing  switch  (2).  Now  the  spray 
system  will  be  filled  automatically 
and  a  strong  spray  stream  will  be 
blowed out of the hand piece for the 
first  15  sec.  Then  you  can  adjust 

the  requested  spray  stream  while 
turning knob (1). 
 
When  turning  off  the  unit  or  the 
spray  function,  the  spray  liquid  is 
pumped  back  completely  into  the 
tank  from  the  hoses. With  this  pro-
cess,  a  bubbling  can  be  heard  in 
the tank. 

 
 

Drying / changing tool 

Press  key  (7)  and  the  hand  piece 
and  spray  will  be  turned  off.  Now 
only  air  will  come  out  of  the  top  of 
the  hand  piece,  which  you  can  use 
to  dry  the  spot  of  the  recent  treat-
ment.  When  you  push  button  (7) 
again, the spray and hand piece are 
working again. 

 

 

This function can also be used 
to change your tools while the 
hand piece is not running. 

 
 

Summary of Contents for Hyperios4 CHROME

Page 1: ...Manual Hyperios4 CHROME Art No 0584 FB 04 293 Rev 0 Date 26 01 2018 ...

Page 2: ...peration steps 11 Work positioning 12 Infilling the spray liquid 13 Insert the bur 14 Adjusting the speed 14 Memory buttons m1 m2 14 Spray 14 Drying changing tool 14 Foot switch 15 Maintenance Care 16 Washer 16 Water filter 16 Cleaning function 17 Cleaning the chuck 17 Maintenance overview 17 Handling Instructions 18 Troubleshooting 19 Technical Details 20 EMC 23 ...

Page 3: ... you give this product to someone else please include this document as it includes important information on how to correctly use the product Short description of functions The Hyperios4 CHROME is a medical device for the treatment of callus and nails The hand piece is compatible with rotating tools such as diamond polishers steel cutters and ceramic cutters These can reach at speeds up to 40 000 r...

Page 4: ...f the hand piece rests secured in the hand piece holder or if you hold it safely in your hand Prevent the permeation of liquids into the control unit The electronic elements could be damaged Clean the outside of the box only with a damped cloth Be sure you unplug the unit before Only when pulling the mains cable you can ensure that the device will be completely disconnected from mains Check the ma...

Page 5: ...se distilled water plus a sterilizing additive e g hadewe AquaHy art no 8555 If you use other brands pay attention to their manuals if you can use them for this purpose Never use tap water or alcoholic fluid or other burnable liquids In case you are working with an alcoholic fluid or other burnable liquids there is always the danger of fire or explosion hazards Never refill the tank directly over ...

Page 6: ...ximum Now start the spray function and the water will spray for approx 3 min The fluid will be blown out of the complete spray system and hand piece Always follow the operating instructions Do not carry out any repairs modifications or maintenance work yourself This is only to be accomplished by an authorized specialist The mains cable may also only be changed by the manufacturer or an authorized ...

Page 7: ...on please adhere to the cleaning and disinfection instructions provided below After every treatment please carry out these steps Remove the rotating tool e g bur first To do so pay attention to the instructions of its manufacturer Remove clothing remains through wiping off easily the unit with a moistened paper cloth Afterwards clean all surfaces with a disinfec tion cloth e g Mikrozid AF cloths o...

Page 8: ...tion 6 changeover switch for clockwise or anticlockwise operation 7 turn off hand piece spray turn on airflow 8 pilot lamp change the washer page 16 9 confirms changing the washer 10 hand piece light ON OFF 11 memory button m1 for speed 12 memory button m2 for speed 13 pilot lamp shines in standby mode 14 ON OFF switch standby 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...

Page 9: ...g trimming and polishing Attention The spray must be in operation during foot care Manicure Grinding milling and polishing works The devise is determined of operating rotating instruments which will not penetrate the skin or murcous membrane if used appropriate but could evoke damages accidently Caution In the area of pedicure you have to work with spray techniques Operation The working area must ...

Page 10: ...plays this symbol Make sure that the tank is filled with spray liquid Ensure that the tool is fixed On the back of the device you will find the main power switch Press this switch to turn the unit on Take up the hand piece To use the hand piece and spray function press the button 14 on the front of the unit ...

Page 11: ...rrelevant for diamond cutters and high grade corundum sanders in terms of effi ciency Working anti clockwise can be beneficial if for example the dust obscures your field of vision when working clockwise Now you can turn on the spray by pushing button 2 By using the adjusting knob 1 you can set the amount of the spray liquid After setting the speed and the spray amount you can start with working I...

Page 12: ... Application Work positioning The following images are examples of how to work on various parts of the foot Always make sure that the foot is firmly gripped and that you can work freely and without strain ...

Page 13: ...build up Put the tank back in the housing Switch on the power switch in the rear of the unit Attention Never use tap water danger of calcifica tion Also do not use explo sive or combustible liquids e g alcohol Note Never refill the tank directly over the unit to avoid in case you acci dently spill water this will flow into the unit The water filter art no 3479 has to be fully under water otherwise...

Page 14: ...ed level simply press button m1 again Further speeds can be saved with m2 Spray Turn on the spray function while pushing switch 2 Now the spray system will be filled automatically and a strong spray stream will be blowed out of the hand piece for the first 15 sec Then you can adjust the requested spray stream while turning knob 1 When turning off the unit or the spray function the spray liquid is ...

Page 15: ...tch Press foot switch long time Switch device ON OFF Press foot switch short time While the spray is on and you push the switch once the spray stream will stop If you push it a second time the hand piece will stop Push ing the switch a third time both functions will start to work again When working with a foot switch take care that you only use instru ments that are approved for use at the maximum...

Page 16: ...s If you set the tool on the LED they may break Push the new washer in the top Now press button 9 to make the LED 8 go out Water Filter Even smallest parts of dirt could block the nozzle which sits in the top Avoiding such blockages there is a water filter installed in the cap of the tank We recommend to change this every time when it is covered with some dirt at least twice a year Still make sure...

Page 17: ... whole spray system Parts of dirt are rinsed into the tank Do not use this water again but clean the tank immediately after realizing a block age in the nozzle Otherwise the dirt parts might stick together Clean the water filter carefully and put it back in the adapter Refill the tank with new water Use the cleaning function only if the nozzle is blocked Do not use the function if the unit works p...

Page 18: ...leaning and disinfection liquids This extra neous rust affects internal parts of the hand piece in form of rust bloom and causes pitting Therefore when you not working with the hand piece always remove the bur tool from it Always make sure that you do not let the hand piece fall down or exposes it to any shocks The ball bearings and the motor could be damaged by this Only work with tools which are...

Page 19: ... the unit Footswitch is used constantly or faulty Pull out the socket of the foot switch to see if this is the cause Pull out the mains plug to reset the device After lengthy use unit does not spray Unit is overheated Slots of the unit should not be covered or blocked Let cool down the unit for a while The unit shows different sound levels at different sites of use The sound level varies according...

Page 20: ... device is designed for extend ed usage Ambient conditions The device is intended for use in dry closed rooms Operation Temp 10 C to 26 C Rel humidity 25 to 75 not condensed Air pressure 700 hPa to 1060 hPa Storage Temp 5 C to 55 C Rel humidity 10 to 95 not condensed Air pressure 500 hPa to 1060 hPa Transport up to 4 weeks Temp 5 C to 55 C Rel humidity 10 to 95 not condensed Air pressure 500 hPa t...

Page 21: ...rent direct current safety fuse type B applied part socket foot switch date of manufacture Disposal The device and filter may contain infectious material such as residual sanding dust or skin particles Therefore please bear in mind the following disposal instructions Disposing of the device Old devices bought in Germany that correspond to the electronic device law should be sent directly to the ma...

Page 22: ...etachable power connection The protective ground conductor is not connected to touchable conduc tive parts Necessary function inspection speed display on off spray clamp ing system Inspection before use During the production control the following inspections as per DIN EN 62353 VDE 0751 1 are conducted among others Optical control unit discharge cur rent 1000 µA discharge current at applied part 1...

Page 23: ... to CISPR 11 Group 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions according to CISPR 11 Class B The device is suitable for use in all establishments including domestic estab lishments and those directly connected to the public low voltage power supply network...

Page 24: ...lity should be that of a typical commercial and or hospital environment Surge accord ing to IEC 61000 4 5 1 kV differen tial mode 2 kV common mode Not applicable Mains power quality should be that of a typical commercial and or hospital environment Voltage dips short interrup tions and voltage varia tions on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for cycle 40 UT 60 dip in UT for...

Page 25: ... separation distance calculated from the equation appropriate for the frequen cy of the transmitter Recommended Separation Distance d 1 17 P d 1 17 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF tra...

Page 26: ...h radiated RF dis turbances are controlled The customer or the user of the device can help prevent electromag netic interference by maintaining a minimum distance between portable and mobile RF com munications equipment transmitters and the pa on Parometer as recommended below ac cording to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separat...

Page 27: ......

Page 28: ...hadewe GmbH Grambartstraße 10 30165 Hannover Germany info hadewe de www hadewe de ...

Reviews: