11 

Technical Data 

art. 6231, 6232, 6233, 6234 

 

Handpiece  

weight: 160 g 
length of handpiece: 126 mm 
handpiece Ø: 13 mm 
motor Ø: 27 mm 
speed: 4.000 

 20.000 rpm 

 
Steering unit 

weight: 0,9 kg 
dimensions:W114 x H86 x D183mm 
input: 230 V ~, 50 Hz 
output: 50 W 
fuse primary: 

F1: 0,315A time delayed safety 
fuse 

fuse secondary: thermal fuse 
protection class II 
 
 

Declaration of Conformity 

We  declare  under  our  sole 
responsibility,  that  the  products  g-
file  6231,  6232,  6233  and  6234  to 
which  this  declaration  relates,  are 
conform  to  the  following  standards 
or normative documents: 
 
2014/30/EU directive EMC 
2014/35/EU low voltage directive  

 

hadewe  GmbH 

  Grambartstraße  10 

30165 Hannover 

 Germany 

 
 

Symbols

 

 

 

 

Caution!  

 

 

Important note! 

 

 

Follow instructions for use! 

 

protection class II 
 

  direct current 

 

  safety fuse 

 

~

 

alternating current 
 

 

date of manufacture, 
name and address 
of the manufacturer

 

 
 

Disposal 

For  disposal  of  this  unit,  please 
contact  the  site,  where  you  bought 
this hadewe product. 
 
 

Safety Check 

A  safety  check  must  be  done  in 
compliance  with  the  regulations  of 
the  country  where  the  device  is 
used.  The  inspection  interval  is 
once a year.  

 
 

Summary of Contents for g-file

Page 1: ...art 6231 6232 6233 6234 FB 04 279 Rev 0 Date 11 02 2019 Manual g file ...

Page 2: ...2 Index of Contents Safety Instructions 3 Application 7 Handling Instructions 8 Troubleshooting 9 Exploded Drawing 10 Technical Data 11 ...

Page 3: ...nd the hazards involved Children should be supervised not to play with the appliance The installation of the working area has to correspond to the relevant regulations Only switch on the unit if the handpiece rests secured in the handpiece holder or if you hold it safely in your hand Prevent any liquids from entering into the control unit since this could damage the electronic elements Clean the o...

Page 4: ...r comes off during the treatment itself the motor blocks during the treatment Turn on the unit only when you hold the handpiece safely in your hand or if it is placed in the handpiece holder Ensure that the patient s foot and the handpiece are held firmly in place during the operation To prevent injury work with extreme caution on the foot of the patient and use only light pressure To ensure that ...

Page 5: ... max speed see manufacturer s instructions or only use the speed required by the instrument the tool is absolutely dry After each use and before each new customer The tools are to be cleaned disinfected and sterilized Caution Handpiece and motor may not be sterilized All surfaces of the handpiece are disinfected Take care that no liquids enter the handpiece Always follow the operating instructions...

Page 6: ...especially not while the equipment is plugged into the socket since some components are under electric voltage Any unauthorized or improper handling e g trying to repair the unit of the equipment will immediately nullify the guarantee warranty and any other liabilities of the manufacturer ...

Page 7: ... ring B clockwise until it snaps in Now the clamp is open Insert the bur all the way to the stop Turn ring B counter clockwise until it snaps in Check that the bur is clamped correctly and cannot be pulled out You may never turn on the unit while the chuck is open This may cause damages at the motor and the electronic For security reasons only change the tool if the handpiece is turned off Adjusti...

Page 8: ...t with chemical cleaning and disinfection liquids This extraneous rust affects internal parts of the handpiece in form of rust bloom and causes pitting Therefore when you not working with the handpiece always remove the bur tool from it Always make sure that you do not let the handpiece fall down or exposes it to any shocks The ball bearings and the motor could be damaged by this Only work with to...

Page 9: ...in a new carbon brush Check the position of the spring on the brush It must be in the middle of the brush The carbon brush is a wearing part and has to be exchanged when worn You will find a pair of carbon brushes enclosed Check the second brush as well When the handpiece becomes warm Should the handpiece become warm or make loud running noises probably the ball bearings in the handpiece front par...

Page 10: ...10 Exploded Drawing Explosive Drawing for Service Repair Shops ...

Page 11: ...6234 to which this declaration relates are conform to the following standards or normative documents 2014 30 EU directive EMC 2014 35 EU low voltage directive hadewe GmbH Grambartstraße 10 30165 Hannover Germany Symbols Caution Important note Follow instructions for use protection class II direct current safety fuse alternating current date of manufacture name and address of the manufacturer Dispo...

Page 12: ...hadewe GmbH Grambartstraße 10 30165 Hannover Germany info hadewe de www hadewe de ...