6
Safety Instructions
Ensure that
the patient’
s foot and the handpiece
are held firmly in place during the operation. To
prevent injury, work with extreme caution on the
foot of the patient and use only light pressure.
If water leaks from the unit, then a disturbance
is present. The unit has to be separated from
the socket and cannot be used any longer.
Only use distilled water plus a sterilizing
additive, e.g. hadewe AquaHy (art. no. 8555). If
you use other brands, pay attention to their
manuals if you can use them for this purpose.
Never use tap water or alcoholic fluid, or other
burnable liquids. In case, you are working with
an alcoholic fluid, or other burnable liquids,
there is always the danger of fire or explosion
hazards. Never refill the tank directly over the
unit to avoid in case you accidently spill water,
this will flow into the unit.
Change the spray liquid every day. Always add
disinfection additive to the distilled water.
Rinse the tank daily and fill it with fresh water
every day.