15 

Application 

Infilling the spray liquid 

 

Switch  off  the  device.  Pull  out  the 
spray  tank  on  its  handle.  Hold  the 
cover  of  the  tank  in  hand  and  un-
screw the tank.  
Now  fill  it  with  distilled  water  and 
anti germ liquid (e.g. AQUAHY item 
no. 8555). 
 

Do  not  exceed  the  maxi-
mum  level  when  filling 
the tank. 

 

Now turn the tank cover back on the 
tank. While screwing back the cover 
on  the  tank,  make  sure  that  is 
closed  correct.  Otherwise  there  will 
no pressure be build up.  
Put the tank back in the housing. 
Switch  on  the  power  switch  in  the 
rear of the unit.   

 

Attention: 

Never  use  tap 

water.  (Danger  of  calcifi-
cation).  Also  do  not  use 

explosive or combustible liquids e.g. 
alcohol.  

Note: 

The filter in the tank (item no. 

3479)  should  be  fully  under  pres-
sure;  otherwise  no  water  will  be 
sucked in. 
 

The  water  filter  has  to  be 
fully under water, otherwise 
no water will be sucked in. 

 

Mounting/Changing the  
filter 

 

Remove  cover  plate  with  filter 
(Art.5115).  Now  the  filter  can  be 
pulled  from  the  plate  and  be 
changed.  Never  work  with  full  filter 
or  without  filter.  Put  up  cover  plate. 
The  cover  plate  must  lie  flat  but 
may move a little. It fits tightly when 
the suction cleaner is in operation. 
 
As  an  option  you  may  also  use  the 
micro  filter  (Art.  No.  5179)  instead 
of paper filter 5115.  
 

 

 

Insert the bur 

 

The  handpiece  is  equipped  with  a 
chuck which can hold all burs with a 
diameter  of  2.35  mm  and  usually 
are used in the fields of podiatry. To 
open  the  chuck,  push  the  black 
knob  forward  with  your  thumb  and 
leave it there. Now insert the tool as 
far  as  possible.  Pull  back  the  black 
knob and the tool will be clamped. 

Summary of Contents for Argos 2

Page 1: ...Art No 0572 FB 04 278 Rev 0 Date 10 05 2016 Manual Argos 0572...

Page 2: ...tion 12 Operation 12 Work positioning 13 Infilling the spray liquid 15 Mounting Changing the filter 15 Insert the bur 15 Adjusting the speed 16 Spray 16 Drying changing tool 16 Footswitch 16 Maintance...

Page 3: ...duct to someone else please include this document as it includes important information on how to correctly use the product Short description of functions The Argos is a device for the treatment of cal...

Page 4: ...lways be easily accessible Only switch on the unit if the handpiece rests secured in the handpiece holder or if you hold it safely in your hand Prevent the permeation of liquids into the control unit...

Page 5: ...should be exchanged at least after 50 operating hours Never use the suction to vacuum any liquids or air water fog which contains particles of liquids If water leaks from the unit then a disturbance...

Page 6: ...ent injury work with extreme caution on the foot of the patient and use only light pressure To ensure that the burr tool sits firmly in the chuck do not work with tools which have an oily shaft are wo...

Page 7: ...orkshops and trained technicians upon re quest Never remove the housing while the equipment is plugged in as some components are live A spare parts list may be sup plied to qualified persons repair wo...

Page 8: ...tion please adhere to the cleaning and disinfection instructions provided be low After every treatment please carry out these steps Remove the rotating tool e g bur first To do so pay attention to the...

Page 9: ...the water tank 6 Water tank 7 ON OFF button line voltage is not disconnected unit in standby 8 handpiece speed down 9 handpiece speed up 10 Switch for modus change spray or suction 11 Switch for turni...

Page 10: ...carry case It is suitable for the following scopes Foot care Milling trimming and polishing Attention The suction or spray must be in operation dur ing foot care Manicure Grinding milling and polishin...

Page 11: ...from the patient s foot The unit should be 30 90cm from the floor Image Example of positioning of foot care unit in treatment room Image Bird s eye view of work area If you have a foot switch insert...

Page 12: ...t for diamond cutters and high grade corundum sanders in terms of effi ciency Working anti clockwise can be beneficial if for example the dust obscures your field of vision when working clockwise Now...

Page 13: ...Application Work positioning The following images are examples of how to work on various parts of the foot Always make sure that the foot is firmly gripped and that you can work freely and without str...

Page 14: ...14 Application...

Page 15: ...item no 3479 should be fully under pres sure otherwise no water will be sucked in The water filter has to be fully under water otherwise no water will be sucked in Mounting Changing the filter Remove...

Page 16: ...e heard in the tank Drying Changing Tool Press key 11 and the handpiece and spray will be turned off Now only air will come out of the top of the handpiece which you can use to dry the spot of the rec...

Page 17: ...find some remove such eventually Please pay attention to the way it was installed Push the new washer in the top Water Filter Art no 3497 Even smallest parts of dirt could block the nozzle which sits...

Page 18: ...rts of dirt are rinsed into the tank Do not use this water again but clean the tank immediately after realiz ing a blockage in the nozzle Otherwise the dirt parts might stick together Clean the water...

Page 19: ...a small drop of oil on your thumb and pointer Rub both fingers until you find very little of oil on your fingers Now take the chuck and put it between these two fingers and turn it around There shoul...

Page 20: ...al cleaning and disinfection liquids This extraneous rost affects internal parts of the handpiece in form of rust bloom and causes pitting Therefore when you not working with the handpiece al ways rem...

Page 21: ...in the manual Rubber ring under the knob twists while using it Twist it of 180 It does not spray Pressure cannot be set up in the tank because the lid is not tightened well There are parts of dirt in...

Page 22: ...ters There is air in the spray system Af ter 5 min running this effect will dis appear Acceleration of the spray stream Note To reduce the starting time of the spray turn the spray regulator on Max Af...

Page 23: ...for extended usage Ambient conditions The device is intended for use in dry closed rooms Operation Temp 10 C to 26 C Rel humidity 25 to 75 not condensed Air pressure 700 hPa to 1060 hPa Storage Temp 5...

Page 24: ...ent Direct current Safety fuse Type B applied part Socket foot switch Date of manufacture Disposal The device and filter may contain infectious material such as residual sanding dust or skin particles...

Page 25: ...tion spray clamping system Inspection before use During the production control the following inspections as per DIN EN 62353 VDE 0751 1 are conducted among others Optical control unit discharge cur re...

Page 26: ......

Page 27: ......

Page 28: ...hadewe GmbH Grambartstra e 10 30165 Hannover Germany info hadewe de www hadewe de...

Reviews:

Related manuals for Argos 2