![Hach TU5300 sc Basic User Manual Download Page 30](http://html.mh-extra.com/html/hach/tu5300-sc/tu5300-sc_basic-user-manual_577151030.webp)
Warning
Description
Solution
NOT DRYING
The instrument cannot regulate the
internal humidity.
Replace the desiccant cartridge. Refer to
on page 28.
If the error continues, contact technical support.
Measurements with limited validity are still available.
PUMP
The air pump for the drying circuit
has a malfunction.
Contact technical support. Measurements with
limited validity are still available.
SENS.DRY: FUNC
The air system of the drying system
has a malfunction.
Contact technical support. Measurements are still
available, but the life of the desiccant cartridge is
decreases.
TURB TOO HIGH
The turbidity reading is not within
the calibration range.
Make sure that the calibration range selected is
applicable to the turbidity value of the sample.
WIPER REPLACE
A wiper replacement is due in the
automatic cleaning module.
Replace the wiper in the automatic cleaning module.
Refer to the documentation supplied with the
automatic cleaning module to replace the wiper.
VIAL CLARITY
The vial or vial compartment is dirty.
Clean or dry the vial and the vial compartment.
10.3 Errors
Errors show on the controller display. To see all of the active errors, push
menu
then select
DIAGNOSTICS>TU5x00 sc>ERROR LIST.
Error
Description
Solution
AUTOCHK. NO FUNC The automatic system check does not
complete.
Contact technical support.
CLEANING MODULE
The automatic cleaning module has a
malfunction.
Contact technical support.
EE RSRVD ERR
There is a problem with the internal memory.
Contact technical support.
FLASH FAIL
The internal calibration memory is corrupted.
Contact technical support.
HUMIDITY PCB
There is humidity or water in the instrument.
Contact technical support.
LASER TOO LOW
The laser has a malfunction.
Contact technical support.
MEAS ELECTRONIC
There is a measurement error. There is a
problem in the electronics unit.
Contact technical support.
PROC HEAD OPEN
The process head is in the open position or
the process head detector has a malfunction.
Turn the process head to the closed
position.
TURB TOO HIGH
The turbidity reading is higher than the
measurement range of the instrument
(700 FNU maximum).
Make sure that the turbidity value of the
sample is within the measurement range
of the instrument.
VIAL PRESENT
There is no vial in the vial compartment.
Install a vial in the vial compartment.
VIAL CLARITY
The vial or vial compartment is dirty.
Clean or dry the vial and the vial
compartment.
WATER INGRESS
9
There is water in the instrument.
Immediately stop flow to the instrument.
Disconnect the sensor cable.
The desiccant cartridge can become hot.
Only touch and remove the desiccant
cartridge when it is at room temperature.
9
Water drops, puddles or runlets that will not damage the instrument may be in the inner of the
enclosure.
30
English
Summary of Contents for TU5300 sc
Page 12: ...12 English ...
Page 15: ...English 15 ...
Page 19: ...English 19 ...
Page 27: ...English 27 ...
Page 41: ...Français 41 ...
Page 44: ...44 Français ...
Page 49: ...Français 49 ...
Page 57: ...Français 57 ...
Page 71: ...Español 71 ...
Page 74: ...74 Español ...
Page 79: ...Español 79 ...
Page 87: ...Español 87 ...
Page 102: ...102 Português ...
Page 105: ...Português 105 ...
Page 110: ...110 Português ...
Page 118: ...118 Português ...
Page 132: ...132 中文 ...
Page 134: ...图 4 检查干燥剂盒 1 安装到期日 mm yyyy 月份和 年份 2 指示灯 淡蓝色 未过期 白色 过期 3 运输安全保护 134 中文 ...
Page 135: ...中文 135 ...
Page 139: ...中文 139 ...
Page 146: ...146 中文 ...
Page 147: ...9 7 更换干燥剂盒 控制器屏幕将显示何时应更换干燥剂盒 请参阅干燥剂盒包装袋中随附的文档更换干燥剂盒 中文 147 ...
Page 159: ...日本語 159 ...
Page 161: ...図 4 乾燥剤カートリッジの検査 1 設置有効期限 mm yyyy 月と 年 2 インジケーター 水色 有効期 限内 白 有効期限切れ 3 輸送安全保護 日本語 161 ...
Page 162: ...162 日本語 ...
Page 167: ...日本語 167 ...
Page 175: ...日本語 175 ...
Page 188: ...188 ไทย ...
Page 191: ...ไทย 191 ...
Page 196: ...196 ไทย ...
Page 204: ...204 ไทย ...
Page 217: ...한글 217 ...
Page 219: ...그림 4 건조제 카트리지 검사 1 설치 날짜 mm yyyy 월 및 연도 2 표시등 연청색 만료되지 않음 흰색 만료됨 3 운송 안전 보호 한글 219 ...
Page 220: ...220 한글 ...
Page 225: ...한글 225 ...
Page 233: ...한글 233 ...
Page 246: ...246 中文 ...
Page 248: ...图 4 檢查乾燥劑濾芯 1 安裝有效期限 mm yyyy 月份 和年份 2 指標 淺藍色 未過期 白色 過期 3 運輸安全防護 248 中文 ...
Page 249: ...中文 249 ...
Page 253: ...中文 253 ...
Page 260: ...260 中文 ...
Page 261: ...9 7 更換乾燥劑濾芯 控制器顯示幕會顯示乾燥劑濾芯的更換到期日 請參閱乾燥劑濾芯袋中隨附的文件 瞭解如何更換乾 燥劑濾芯 中文 261 ...
Page 264: ......
Page 265: ......