5.5 Replace the wiper
To make sure that the vial is cleaned fully, replace the wiper periodically.
1.
Push
menu
.
2.
Select SENSOR SETUP>[select analyzer]>DIAG/TEST>MAINTENANCE>WIPER REPLACE.
3.
Stop the sample flow.
4.
Remove the cleaning module.
5.
Remove the vial. Refer to steps 1 to 5 of
6.
Complete the steps that show on the controller display. Install the vial wiper (silicone or fiber) that
is applicable to the sample type. Refer to the illustrated steps that follow.
The date that the wiper was replaced is automatically saved after the last screen shows.
7.
Install the vial. Refer to steps 8 to 12 of
on page 14.
English
17
Summary of Contents for TU5 Series
Page 2: ...Table of Contents English 3 Français 19 Español 36 Português 53 中文 70 日本語 86 한글 103 ไทย 119 2 ...
Page 9: ...English 9 ...
Page 10: ...10 English ...
Page 11: ...English 11 ...
Page 16: ...16 English ...
Page 25: ...Français 25 ...
Page 26: ...26 Français ...
Page 27: ...Français 27 ...
Page 32: ...32 Français ...
Page 42: ...42 Español ...
Page 43: ...Español 43 ...
Page 44: ...44 Español ...
Page 49: ...Español 49 ...
Page 50: ...50 Español ...
Page 59: ...Português 59 ...
Page 60: ...60 Português ...
Page 61: ...Português 61 ...
Page 66: ...66 Português ...
Page 76: ...76 中文 ...
Page 77: ...中文 77 ...
Page 78: ...78 中文 ...
Page 83: ...中文 83 ...
Page 92: ...92 日本語 ...
Page 93: ...日本語 93 ...
Page 94: ...94 日本語 ...
Page 99: ...日本語 99 ...
Page 109: ...한글 109 ...
Page 110: ...110 한글 ...
Page 111: ...한글 111 ...
Page 116: ...116 한글 ...
Page 125: ...ไทย 125 ...
Page 126: ...126 ไทย ...
Page 127: ...ไทย 127 ...
Page 132: ...132 ไทย ...
Page 133: ...ไทย 133 ...
Page 136: ......
Page 137: ......