Section 1 Specifications
Specifications are subject to change without notice.
Specification
Details
Measurement method
Nephelometric
Regulatory
Meets ISO 7027, DIN EN 27027 and DIN 38404
ASTM D7315 - Standard Test Method for Determination of Turbidity
Above 1 Turbidity Unit (TU) in Static Mode
ASTM D6855 - Standard Test Method for Determination of Turbidity
Below 5 NTU in Static Mode
Dimensions (W x D x H) 39.5 x 30.5 x 15.3 cm (15.6 x 12.0 x 6.02 in.)
Weight
2.9 kg (6.4 lb)
Enclosure
IP30; indoor use only
Protection Class
External power supply: Protection Class I; instrument: Protection Class II
Pollution degree
2
Installation category
External power supply: Category II; instrument: Category I
Power requirements
Instrument: 12 VDC, 3.4 A; power supply: 100–240 VAC, 50/60 Hz
Operating temperature
0 to 40 °C (32 to 104 °F)
Storage temperature
–20 to 60 °C (–4 to 140 °F)
Humidity
5 to 95% relative humidity, non-condensing
Display
17.8 mm (7 in.) color touch screen
Light source
Light-emitting diode (LED) at 860 ± 30 nm
Measurement units
FNU, NTU, EBC, Abs (absorbance), %T (% transmittance) and mg/L
(degree)
Range
FNU (Ratio on): 0–1000
FNU (Ratio off): 0–40
NTU (Ratio on): 0–10,000 auto decimal
NTU (Ratio off): 0–40
EBC (Ratio on): 0–2450 auto decimal
EBC (Ratio off): 0–9.8
Absorbance (auto range): 0–2.00
Transmittance (%): 1.0–100
Degree: 0–100 mg/L