![Hach TL2300 Basic User Manual Download Page 21](http://html2.mh-extra.com/html/hach/tl2300/tl2300_basic-user-manual_3423613021.webp)
Instrument utilities
1.
Push
Home
to see the instrument model, version, serial number and location name.
2.
Push
Diagnostics
.
3.
Select an option.
Option
Description
Factory Service
For factory/service use only.
Instrument Backup Store
—Saves a backup of all the instrument settings and log files to a USB flash drive.
Restore
—Copies the instrument settings and log files from a USB flash drive to the
instrument. Overwrites all the instrument settings.
Instrument Update
Installs an instrument update on the instrument from a USB flash drive.
Service Time
Shows the date entered for the last service date and for the next service date. When
set to on, a service reminder shows on the display when service is due.
Troubleshooting
Message
Solution
Startup
The self-check stopped.
Hardware error.
Set the power to off, wait 20 seconds and then set the power to on again. If the
self check is not successful, record the error number and contact technical
support.
Error numbers: 0: RTC; 1: Touch IC; 3: Dark voltage—Close the door until a click
is heard. Start the instrument again. 4: Amplifier coefficient—Make sure that the
power supply is connected to an electrical outlet that has a protective earth
ground. 7: Lamp voltage—Make sure that the correct power supply is used. 8:
Transmission voltage drift—If the lamp was replaced, calibrate the instrument. If a
vial was in the sample compartment during the self-test at startup, remove the vial.
9: SDRAM; 10: NOR flash; 11: SPI flash; 12: Battery voltage; 13: Power supply
voltage—Make sure that the correct power supply is used.
Next calibration is due!
Calibrate the instrument. Refer to
Calibrate the turbidimeter with StablCal
on page 13.
Note: The calibration reminder is set to on. Refer to
Configure the calibration settings
on page 13.
Next service is due!
Contact technical support.
Note: The service reminder is set to on. Refer to
Next verification is due!
Do a calibration verification. Refer to the expanded user manual on the
manufacturer's website.
Note: The verification reminder is set to on.
Reading
Hardware error / instrument
error
Set the power to off, wait 20 seconds and then set the power to on again. If the
problem continues, contact technical support.
The calibration range is
exceeded.
The measured turbidity is more than the calibration range of the instrument. Select
a calibration curve for the full measurement range. Refer to
on page 13.
The measurement range is
exceeded.
The measured turbidity is more than the measurement range of the instrument.
Calibration/Verification
Instrument error
Examine the standards. Start the calibration or verification again.
If calibration (or verification) is not successful, contact technical support.
English
21
Summary of Contents for TL2300
Page 2: ...English 3 Fran ais 24 Espa ol 47 Portugu s 69 92 110 131 152 2...
Page 95: ...A TL2300 90 USB USB 1 1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 95...
Page 97: ...3 3 1 4 7 2 8 1 3 9 4 10 5 11 6 12 1 4 97...
Page 98: ...1 105 98 1 2 USB 98...
Page 103: ...13 14 StablCal 7 13 NTU 15 16 Gelex StablCal 1 ID 3 2 3 4 Gelex 5 6 7 8 103...
Page 104: ...USEPA NPDWR NPDES USEPA 30 60 24 ID ID 100 99 1 ID 3 2 3 ID ID ID 4 30 mL 104...
Page 105: ...5 6 7 8 9 10 1 2 3 4 ID ID ID ID USB USB 2 0 bmp csv xml 105...
Page 106: ...1 USB USB 2 3 4 5 105 6 1 2 105 3 1 2 1 2 3 106...
Page 109: ...99 USB dd mm yyyy hh mm dd mm yyyy hh mm 109...
Page 112: ...EN 55011 CISPR 11 A IECS 003 A A FCC PART 15 A FCC 15 1 2 FCC 15 A 112...
Page 113: ...1 2 3 4 5 EMC A A A TL2300 90 1 USB USB 1 113...
Page 114: ...1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 114...
Page 117: ...1 126 118 117...
Page 124: ...USEPA 1 NPDWR NPDES USEPA 30 60 1 24 1 124...
Page 127: ...1 2 1 2 3 30 60 127...
Page 130: ...Auto Send 118 ID USB ID dd mm yyyy hh mm ID Web dd mm yyyy hh mm ID ID ID 130...
Page 134: ...EMC A A A TL2300 90 USB USB 1 134...
Page 135: ...1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 135...
Page 137: ...3 3 1 ID 4 7 2 8 1 3 9 4 10 5 11 6 12 1 ID ID ID ID ID 4 137...
Page 138: ...1 146 139 138...
Page 146: ...5 6 7 8 9 10 Done 1 2 3 Filter Filter Settings 4 146...
Page 148: ...1 2 3 30 60 1 2 3 USB USB USB 148...
Page 149: ...20 0 RTC 1 IC 3 4 7 8 9 SDRAM 10 NOR 11 SPI 12 13 StablCal 141 141 148 20 141 StablCal 141 149...
Page 151: ...dd mm yyyy hh mm ID ID ID 151...
Page 156: ...1 1 6 2 EPA 7 3 8 USB 4 9 5 10 USB 2 156...
Page 158: ...3 3 1 ID 4 7 2 8 1 3 9 4 10 5 11 6 12 1 Login ID Login Logout ID ID Sample ID ID ID 4 158...
Page 159: ...1 Verification Data Log 167 Setup 160 Diagnostics Timer 159...
Page 173: ......