Option
Description
Options
Sets the following options:
•
Set the application parameters view to expert or basic mode
•
When the instrument is set to off, sets the syringe to empty into the
titrant bottle
•
Changes the temperature display unit
•
Prints the measurement and derivative curves if a printer is connected
•
Specify if a balance is connected
•
Specify if an external propeller stirrer is connected.
Note: When the instrument is set to on, the application parameters view is
always set to basic mode. Changes made to application data with the
instrument in expert mode are kept.
Security
Change the password. Specify which of these options are password
protected:
•
Settings (all items except Info)
•
Maintenance schedule and summary
•
Delete data log
•
Export data log
•
Electrode calibration
•
Titrant calibration
•
Sample volume calibration
•
Syringe verification test
3.
Push
Back
.
6.2 Install the applications
Use the supplied USB key to install the applications. The instrument can install a
maximum of five applications for each syringe installed. For two syringes, the installed
applications shown on the top line of the display refer to syringe one and the installed
applications shown on the second line refer to syringe two.
If any errors occur during installation, refer to
on page 49.
1.
Push
Home
to go to the main menu.
2.
Connect the USB key to the USB port on the side of the instrument. The applications
on the USB key show on the display.
3.
Push the arrow keys to highlight and select an application to install. Push the left or
right arrow key to select it. Do this step again to select additional applications to
install.
4.
Push
Import
to install the selected applications.
5.
Push
OK
to complete the installation. The installed applications show on the home
screen.
Note: To install more applications, push Home to go to the home screen, then remove the
USB key and reconnect it.
6.3 Prepare the instrument for measurement
1.
From the home screen, select
Purge
. All attached devices are listed.
2.
Select
All elements
to purge all the attached devices, or select one device to purge.
Push
Select
. Air is removed from the device and filled with liquid from the bottle.
3.
Push
OK
when the operation has completed.
4.
Make sure that there are no air bubbles in the device. Do step
air bubbles.
Startup
30
Summary of Contents for TitraLab AT1102
Page 1: ...DOC022 52 93074 TitraLab AT1000 series workstations User Manual 03 2019 Edition 7...
Page 2: ......
Page 6: ...Specifications 4...
Page 20: ...Installation 18...
Page 25: ...Install accessories 23...
Page 26: ...Install accessories 24...
Page 34: ...Startup 32...
Page 42: ...Standard operations 40...
Page 50: ...Maintenance 48...
Page 58: ...Troubleshooting 56...
Page 64: ...Replacement parts and accessories 62...
Page 74: ...Change application parameters 72...
Page 79: ......