![Hach SP-510 User Manual Download Page 13](http://html2.mh-extra.com/html/hach/sp-510/sp-510_user-manual_3150143013.webp)
Figure 5 Sampling methods
1
Air
2
Sample flow
Install the pump valve pressure plate
The pressure plate and screws are supplied with the installation kit. Remove the tape from the pump
tubes before installation.
Make sure to turn the screws in small increments and move from one screw to the other so that the
plate is pulled down equally. Do not overtighten. To install the pressure plate, refer to
Figure 6
.
Figure 6 Install the pressure plate
Install the sample conditioning kit
The sample conditioning kit is supplied with the instrument. The kit removes large particles from the
sample stream with a 40-mesh strainer. The ball valve on the raw sample inlet controls the flow to
the filter. The ball valve on the instrument supply line controls the flow rate of filtered sample to the
instrument. For complete sample conditioning installation instructions, refer to the installation kit
documentation. Refer to
Figure 7
.
English
11
Summary of Contents for SP-510
Page 1: ...DOC023 53 80422 SP 510 01 2017 Edition 8 User Manual...
Page 2: ......
Page 28: ......
Page 29: ......