Prepare for storage
Before short-term storage (three hours or less), flush the flow cell assembly and tubing. Refer to
Flush the flow cell assembly and tubing
on page 11. Keep the flow cell assembly full of distilled or
deionized water to prevent air locks and the collection of particles on the parts.
Before long-term storage, remove and clean the flow cell assembly. Do steps
of
on page 12. Keep the flow cell assembly full of distilled or deionized water to prevent
air locks and the collection of particles on the parts.
Maintenance
Clean the instrument
Clean the exterior of the instrument with a moist cloth and a mild soap solution and then wipe the
instrument dry as necessary.
Clean the flow cell assembly
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
N O T I C E
Do not to touch or scratch the flow cell. Contamination or scratches on the glass can cause measurement errors.
Clean the flow cell assembly before initial use and as necessary to remove contamination.
Items supplied by the user:
Laboratory glass-cleaning
detergent
Laboratory
detergent
Distilled or deionized
water
Soft, lint-free cloth
Silicone oil
Oiling cloth
1:1 Hydrochloric acid
1
1.
Remove and disassemble the flow cell assembly. Refer to
2.
Clean the glass parts as follows:
a.
Clean the internal and external surfaces of the glass parts with a laboratory glass-cleaning
detergent.
b.
Fully rinse the glass parts many times with distilled or deionized water.
Note: Steam sterilize the tubing, flow cells, and caps in the flow cell assembly if necessary.
c.
For low-range turbidity samples or dilution water measurements, clean the internal and
external surfaces of the glass parts with 1:1 hydrochloric acid.
1
Only necessary for low-range turbidity measurements
12
English
Summary of Contents for SIP 10 kit
Page 2: ...English 3 Fran ais 17 Espa ol 31 Portugu s 45 59 72 85 98 2...
Page 8: ...8 English...
Page 9: ...English 9...
Page 22: ...22 Fran ais...
Page 23: ...Fran ais 23...
Page 29: ...Fran ais 29...
Page 36: ...36 Espa ol...
Page 37: ...Espa ol 37...
Page 50: ...50 Portugu s...
Page 51: ...Portugu s 51...
Page 61: ...TL23xx SIP 10 TL23xx 1 61...
Page 62: ...1 1 USB 4 7 2 TL23xx 5 8 SIP 10 3 6 9 2 62...
Page 64: ...64...
Page 65: ...65...
Page 66: ...USB USB 1 2 3 66...
Page 68: ...MSDS SDS 1 1 1 1 3 2 a b c 1 1 3 4 5 a b c 1 68...
Page 69: ...d 6 3 3 O O O 2 63 67 Lagoprene 67 69...
Page 70: ...70...
Page 74: ...1 2 3 4 5 TL23xx SIP 10 TL23xx 1 74...
Page 75: ...1 1 USB 4 7 2 TL23xx 5 8 SIP 10 3 6 9 2 75...
Page 77: ...77...
Page 78: ...78...
Page 79: ...USB USB 1 2 3 79...
Page 81: ...81 1 4 MSDS SDS 1 1 1 1 3 2 a b c 1 1 3 4 1 81...
Page 82: ...5 a b c 1 2 d 6 3 3 O O O 2 76 80 Lagoprene 80 82...
Page 83: ...83...
Page 87: ...1 2 3 4 5 TL23xx SIP 10 TL23xx 1 87...
Page 88: ...1 1 USB 4 7 2 TL23xx 5 8 SIP 10 3 6 9 2 88...
Page 90: ...90...
Page 91: ...91...
Page 92: ...USB USB 1 2 3 92...
Page 94: ...94 1 4 MSDS SDS 1 1 1 1 3 2 a b c 1 1 3 4 1 94...
Page 95: ...5 a b c d 6 3 3 O O O 2 89 93 Lagoprene 93 95...
Page 96: ...96...
Page 100: ...Class A Part 15 FCC Rules 1 2 3 4 5 SIP 10 TL23xx TL23xx 1 100...
Page 101: ...1 1 USB 4 7 2 TL23xx 5 8 SIP 10 3 6 9 2 101...
Page 103: ...103...
Page 104: ...104...
Page 105: ...USB USB 1 2 Power 105...
Page 107: ...1 4 107 MSDS SDS 1 11 1 3 2 a b c 1 1 3 4 1 107...
Page 108: ...5 a b c d 6 3 3 2 102 106 Lagoprene 106 108...
Page 109: ...109...
Page 111: ......