![Hach SIP 10 kit User Instructions Download Page 11](http://html1.mh-extra.com/html/hach/sip-10-kit/sip-10-kit_user-instructions_3258269011.webp)
4.
Select On to enable the sipper operation.
Note: When On is selected, the sipper unit completes a sipper cycle before each measurement and a purge
cycle (if enabled) after each measurement.
5.
Select and configure each option.
Option
Description
Sip Time
Sets the period of time that the sipper unit pulls the sample through the flow cell before each
measurement. Options: 1 to 99 seconds (default: 8 seconds). The pump flow speed is
1 mL/sec.
Settle Time
Sets the period of time that the sample sits in the flow cell before it is measured. Options: 1 to
999 seconds (default: 5 seconds). Use the Settle Time setting to let the air bubbles and
turbulence in the sample decrease before each measurement.
Purge Time
Sets the period of time that the sipper unit pulls a user-supplied rinse agent through the flow
cell when a purge cycle is done. Options: Off (default) or 1 to 99 seconds (default: 8 seconds).
Note: A minimum of 120 mL of rinse agent is necessary to remove sample from the tubing.
Purge Start
Sets the purge cycle to start automatically after a measurement or when the user pushes
Purge. Options: Auto (default) or Manual.
Note: The Purge Start button is only enabled when Purge Time is set to on.
6.
Push
OK
.
To exit and not save the changes, push
Cancel
. To change the settings to the factory default
settings, push
Default
.
Flush the flow cell assembly and tubing
Before initial use of a new flow cell assembly or tubing, flush the flow cell assembly and tubing.
1.
Put the inlet tubing in approximately 400 mL of deionized water.
Make sure that the tube opening is at the bottom of the container.
2.
Push
Sipper
, then push
Purge
to start a purge cycle.
3.
Do step
again until approximately 360 mL of deonized water is pulled through the flow cell
assembly and tubing.
4.
Push
OK
.
Measure a sample
1.
Do a purge as follows:
a.
Put the inlet tubing in deionized water.
b.
Push
Sipper
, then push
Purge
to start a purge cycle.
c.
Push
OK
.
2.
Put the inlet tubing in the sample.
3.
Push
Read
. The sipper unit completes a sipper cycle, then the turbidimeter measures the
sample.
If the Purge Start setting is set to Auto, a purge cycle is completed after the measurement.
If the Purge Start setting is set to Manual, the Read button changes to a Purge button.
4.
Push
Purge
to start a purge cycle if applicable.
5.
Do steps
and
again until the sample measurements are complete.
6.
Do a purge. Refer to step
English
11
Summary of Contents for SIP 10 kit
Page 2: ...English 3 Fran ais 17 Espa ol 31 Portugu s 45 59 72 85 98 2...
Page 8: ...8 English...
Page 9: ...English 9...
Page 22: ...22 Fran ais...
Page 23: ...Fran ais 23...
Page 29: ...Fran ais 29...
Page 36: ...36 Espa ol...
Page 37: ...Espa ol 37...
Page 50: ...50 Portugu s...
Page 51: ...Portugu s 51...
Page 61: ...TL23xx SIP 10 TL23xx 1 61...
Page 62: ...1 1 USB 4 7 2 TL23xx 5 8 SIP 10 3 6 9 2 62...
Page 64: ...64...
Page 65: ...65...
Page 66: ...USB USB 1 2 3 66...
Page 68: ...MSDS SDS 1 1 1 1 3 2 a b c 1 1 3 4 5 a b c 1 68...
Page 69: ...d 6 3 3 O O O 2 63 67 Lagoprene 67 69...
Page 70: ...70...
Page 74: ...1 2 3 4 5 TL23xx SIP 10 TL23xx 1 74...
Page 75: ...1 1 USB 4 7 2 TL23xx 5 8 SIP 10 3 6 9 2 75...
Page 77: ...77...
Page 78: ...78...
Page 79: ...USB USB 1 2 3 79...
Page 81: ...81 1 4 MSDS SDS 1 1 1 1 3 2 a b c 1 1 3 4 1 81...
Page 82: ...5 a b c 1 2 d 6 3 3 O O O 2 76 80 Lagoprene 80 82...
Page 83: ...83...
Page 87: ...1 2 3 4 5 TL23xx SIP 10 TL23xx 1 87...
Page 88: ...1 1 USB 4 7 2 TL23xx 5 8 SIP 10 3 6 9 2 88...
Page 90: ...90...
Page 91: ...91...
Page 92: ...USB USB 1 2 3 92...
Page 94: ...94 1 4 MSDS SDS 1 1 1 1 3 2 a b c 1 1 3 4 1 94...
Page 95: ...5 a b c d 6 3 3 O O O 2 89 93 Lagoprene 93 95...
Page 96: ...96...
Page 100: ...Class A Part 15 FCC Rules 1 2 3 4 5 SIP 10 TL23xx TL23xx 1 100...
Page 101: ...1 1 USB 4 7 2 TL23xx 5 8 SIP 10 3 6 9 2 101...
Page 103: ...103...
Page 104: ...104...
Page 105: ...USB USB 1 2 Power 105...
Page 107: ...1 4 107 MSDS SDS 1 11 1 3 2 a b c 1 1 3 4 1 107...
Page 108: ...5 a b c d 6 3 3 2 102 106 Lagoprene 106 108...
Page 109: ...109...
Page 111: ......