shows the wiring of the pH probe in the pre-amplifier junction box. Since the pH
reading needs to compensate for temperature variation, a temperature sensor is built into
every pH electrode. The pH probe consists of five wires, three for the pH probe and two
for the temperature sensor.
Stray electrical currents are sometimes found in wastewater streams. These stray
electrical currents can affect the pH readings. In the case of stray electrical currents, a
grounded pH probe is required (
).
Table 8 pH connector pin assignments (J-3)
Pin
Signal Description
Wire Color
A
+5 V dc
White
B
ground
Blue
C
reference
Yellow
D
pH/ORP
Black
E
-5 V dc
Red
F
RTD
Green
Figure 27 pH probe wiring to junction box (un-grounded)
1
Yellow
4
pH
2
Green
5
Glass
3
Clear
6
Red