93
Appendix A Calibration solutions
Note:
Use the tables in this section to calibrate the transmitter when
automatic temperature compensation is set to off.
Table 18 Potassium Chloride solutions (Conductivity in mS/cm)
Temperature
Concentration
1
°C
°F
0.01 mol/L
0.1 mol/L
1 mol/L
0
32
0.776
7.15
65.41
5
41
0.896
8.22
74.14
10
50
1.02
9.33
83.19
15
59
1.147
10.48
92.52
16
60.8
1.173
10.72
94.41
17
62.6
1.199
10.95
96.31
18
64.4
1.225
11.19
98.22
19
66.2
1.251
11.43
100.14
20
68
1.278
11.67
102.07
21
69.8
1.305
11.91
104
22
71.6
1.332
12.15
105.94
23
73.4
1.359
12.39
107.89
24
75.2
1.386
12.64
109.84
25
77
1.413
12.88
111.8
26
78.8
1.441
13.13
113.77
27
80.6
1.468
13.37
115.74
28
82.4
1.496
13.62
29
84.2
1.524
13.87
30
86
1.552
14.12
31
87.8
1.581
14.37
32
89.6
1.609
14.62
33
91.4
1.638
14.88
34
93.2
1.667
15.13
35
95
1.696
15.39
36
96.8
15.64
Summary of Contents for si792 T
Page 2: ......
Page 12: ...Specifications 10...
Page 32: ...Installation 30...
Page 36: ...Interface and navigation 34...
Page 66: ...Operation Foundation Fieldbus 64...
Page 80: ...Maintenance 78...
Page 86: ...Troubleshooting 84...
Page 88: ...Replacement parts 86...
Page 94: ...Warranty liability and complaints 92...
Page 98: ...Calibration solutions 96...
Page 106: ...Concentration curves 104...
Page 112: ...ATEX safety instructions 110 Figure 20 Sanitary style sensor...
Page 116: ...Index 114...