During configuration the transmitter remains in the Hold mode for
safety reasons. The loop current is frozen at the value specified in the
o1.HoLD menu option. The Sensoface icon is inactive. The
configuration mode indicator is displayed (
).
5.2.1
Current output configuration
Select the sensor
o1.SnSR
si792 T:
2208 SIE (7MA2200 series; default)
8398 PM
Other
si792 E:
3700 GLI (default)
Other
Enter the cell factor
1
o1.CELL
0.100–20.000 (3.00 default—si792 T; 4.70 default—si792 E)
Enter the transfer ratio
1
o1.SFC
1.00–200.00 (68.0 default—si792 T; 25.0 default—si792 E)
Select the temperature sensor
1
o1.rTD
Pt100 (default—si792 T)
Pt1000 (default—si792 E for GLI/Hach 3700 series sensors)
NTC100
NTC30
Select the range and units
o1. UnIT
0.000 mS/cm (si792 T)
Conductivity range and resolution
00.00 mS/cm
000.0 mS/cm (default)
0000 mS/cm
0.000 S/m
00.00 S/m
000.0 SAL
Salinity (SAL)
000.0%
Concentration (Conc)