background image

© Hach Company, 2008, 2016. All rights reserved. Printed in Germany.

Edition 2 06/2016

Quick start guide

si792 E/T 2-wire conductivity transmitter 

DOC016.53.00796

4–20 mA signal values

Default settings: 

• 0 mS at 4 mA
• 100 mS at 20 mA

1. 

Press 

CONF

, enter 

1200

 for the “out. 1MNU” menu.

2. 

Press 

ENTER

 to go to the "o1.4mA" or "o1.20mA" sub-menu. 

3. 

Set the values for 4 mA and 20 mA. 

4. 

Press 

CONF

, then 

ENTER

 to exit the configuration mode.

Calibration

Zero calibration in air

1. 

Press 

CAL

, enter passcode: 1001, 

ENTER

. CAL ZRO will be displayed for 3 seconds. 

2. 

Remove and clean the sensor.

Note: 

Make sure that the sensor is dry.

3. 

Use the 

UP ARROW

 and 

RIGHT ARROW

 to change the value until zero is displayed as the 

conductivity value. 

Note: 

If the entry of the zero point value takes longer than 6 seconds, the display will alternate 

between showing the conductivity value and the temperature value.

4. 

Press 

ENTER

 to confirm.

5. 

The display will show the new cell constant value and zero point.

 

Press 

ENTER

.

6. 

Immerse the sensor in the process water.

7. 

The transmitter remains in the HoLD mode. Press 

ENTER

 again.

The transmitter remains in HoLD mode for approximately 20 seconds after calibration and 
returns to the measurement mode.

Note: 

Repeat the calibration when “Err” is displayed.

Calibration with a calibration solution

Use a calibration solution with a known conductivity value to adjust the transmitter to show 
the same value as the solution. 

Note: 

Stabilize the temperature during calibration.

1. 

Press 

CAL

, enter passcode: 0110, 

ENTER

. CAL SOL will be displayed for 3 seconds. The 

transmitter is ready for calibration.

2. 

Remove and clean the sensor.

3. 

Immerse the sensor in the calibration solution. 

4. 

Use the 

UP ARROW

 and 

RIGHT ARROW

 to enter the known value of the calibration solution.

Note: 

If the entry of the calibration solution value takes longer than 6 seconds, the display will 

alternate between showing the cell constant and the temperature value.

5. 

Press 

ENTER

 to save the value.

6. 

The display will show the new cell constant. Press 

ENTER

.

7. 

The transmitter remains in HoLD mode. Press 

ENTER

 again.

The transmitter remains in HoLD mode for approximately 20 seconds after calibration and 
returns to the measurement mode.
The display will show the temperature-corrected value of the calibration solution (unless 
temperature compensation is set to off in the configuration menu).

Note: 

Repeat the calibration when “Err” is displayed.

DANGER
Explosion hazard. This Quick Start Guide is intended for setup of the si792 
transmitter in a non-hazardous location. It is not a substitute for the instructions in 
the user manual (DOC026.53.00796). Installation in a hazardous location must follow 
the applicable control drawing or ATEX certificate (refer to the documentation CD).

DANGER: Electric shock hazard. Disconnect power to all wiring before making 
connection to the si792 E/T series transmitters.

Connect 24 VDC power and communications

Remove the power source before making any electrical connections to the si792 transmitter.

Connect the sensor—Hach GLI 3700 series inductive conductivity sensors

For sensors other than Hach 3700 series, refer to the user manual.

Basic configuration—4–20 mA

1. 

Press 

CONF

 and enter 

1200

 to enter the configuration mode. The following menus can 

be changed:

• 4–20 mA current output—out.1MNU
• Temperature compensation—tc MNU
• Alarms—ALrt MNU 

2. 

To move to a different menu, press the 

RIGHT ARROW

 ">", then press enter.

3. 

To change a setting within a menu, use the 

UP ARROW

 "^" or 

RIGHT ARROW

 ">" buttons 

to change the setting. 

4. 

Press 

ENTER

 to move to the next item in the menu. 

To exit the configuration mode at any time, press 

CONF

 and then 

ENTER

. Refer to the user 

manual for additional configuration options.

Terminal No.

si792 transmitter version

4–20 mA/HART

Profibus PA

Foundation Fieldbus

10

(+)

(–)

(–)

11

(–)

(+)

(+)

Jumper

Terminal No. 3700 series color 7MA2200 series color

1

white

green

Jumper between terminals 2 

and 3

2

blue

yellow

3

yellow

white

4

green

brown

5

shield

violet/black

7

blue/grey

8

red

red/pink

Reviews: