Operation—si792(x) D (4–20 mA/HART)
41
5.2.4
Calibration mode configuration
An overview of the calibration setup menu is shown in
. Refer
for buffer tables.
5.2.5
Alarm settings configuration
An overview of the alarm setup menu is shown in
Table 12 Calibration mode setup menu—CAL
MNU
Select calibration units
CA. MOD
SAt (default)
Conc
Enter calibration timer interval
CA.tiME
Enter the time interval for calibration (0 to 9999 hours; default: 0 h)
Table 13 Output setup menu for alarm settings—ALrt
MNU
Select Sensocheck on or off
AL.SnSO
CHECK ON
Continuous evaluation of sensor function
CHECK OFF (default)
Enter alarm delay
AL.dLY
0010 sec (default)
Range: 0000–0600 sec
LED in Hold mode
AL.LED
HOLD ON
LED blinks during hold
HOLD OFF
LED off during hold
Summary of Contents for si792 D series
Page 2: ...2 ...
Page 6: ...Table of Contents 4 ...
Page 12: ...Specifications 10 ...
Page 36: ...Interface and navigation 34 ...
Page 44: ...Operation si792 x D 4 20 mA HART 42 ...
Page 66: ...Operation Foundation Fieldbus 64 ...
Page 72: ...Operation Profibus PA 70 ...
Page 80: ...Maintenance 78 ...
Page 86: ...Troubleshooting 84 ...
Page 88: ...Parts and accessories 86 ...
Page 93: ...Index 91 Z zero adjustment 72 ...
Page 94: ...Index 92 ...
Page 95: ...93 ...