27
SECTION 3,
continued
Entering the meter altitude
When the meter is received from the factory, the default altitude is
0 meters. To change the value:
1. Access the setup mode by pressing the
SETUP
key.
2. Press the
UP ARROW
five times. A
6
will begin flashing in the
upper right hand corner of the display and the current altitude
entry will be displayed.
3. To change the value, press the
ENTER
key. The number pad
icon will appear in the lower display.
4. Use the number keys to enter the altitude or press the
EXIT
key to leave the value unchanged. Press the
ENTER
key and
then the
EXIT
key when the desired altitude has been entered.
The meter will return to the Read Mode. The displayed dissolved
oxygen in % saturation will be adjusted according to this entry.
3.1.7
Setting the Time
This feature sets the clock in the instrument using military time.
For example, 3:00 p.m. is entered as 15:00. To change the value:
1. Access the setup mode by pressing the
SETUP
key.
2. Press the
UP ARROW
six times. A
7
will begin flashing in the
upper right hand corner of the display. The Time icon will
appear. The current time entry will be displayed.
3. To change the time, press the
ENTER
key. The number pad
icon will appear in the lower display.
4. Press the
EXIT
key to leave the value unchanged. Use the
number keys to enter the time. Press the
ENTER
key, then the
EXIT
key when the desired time has been entered.
The meter will return to the Read Mode.