Change the stirring settings
The magnetic stirrer can be turned on and the stirring speed can be
changed in the Stirring menu.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Stirring and confirm.
3.
To turn on/off the stirrer push .
4.
When the stirrer is turned on, use the or to change the stirring
speed in %.
Note: Use the or to change the stirring speed during calibration and
during a measurement.
Change the temperature units
The temperature units can be changed to Celsius or Fahrenheit.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Temperature units and confirm.
3.
Use the or to select between Celsius or Fahrenheit and confirm.
Data logger
Display data
The Display data log contains Measurement data, Electrode report and
Calibration data. The stored data can be send to a printer or to a PC.
When the data log becomes full (400 data points), the oldest data point
is deleted when a new data point is added.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select DATA LOGGER and confirm.
3.
Use the or to select Display data and confirm.
4.
Use the or to select
Option
Description
Measurement data
Measurement data—stores automatically each time a
sample is measured
Electrode report
Electrode report—stores automatically the electrode
history and measurement conditions
Calibration data
Calibration data—stores automatically the current
calibration
Delete data
The entire measurement data or electrode report log can be deleted to
remove data that has already been sent to a printer or PC.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select DATA LOGGER and confirm.
3.
Use the or to select Erase and confirm.
4.
Use the or to select Measurement data or Electrode report and
confirm. Confirm again to delete the data.
The entire log is deleted at once.
Send data to a printer or to a computer
N O T I C E
The data output (printer or PC) needs to be selected first, so that the Print menu
is available (refer to
Select the data output
on page 11).
Note: Refer to
Report output
on page 13 to select the report output type.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select DATA LOGGER and confirm.
3.
Use the or to select Print and confirm. Select one of the
following options and confirm with to print the data: Measurement
data, Electrode data, Calibration data, Calibration report or
Instrument condit.
12
English