21
10. Standard solutions
The conductivity values of the standard solutions are shown by temperature in Table 1.
Table 1. Stored conductivity and temperature values
Temperature
Conductivity (CE)
°C
°F
µS/cm
µS/cm
mS/cm
mS/cm
15.0 59
119
1147 10.48
92.5
16.0 60.8
122
1173
10.72
94.4
17.0 62.6
125
1199
10.95
96.3
18.0 64.4
127
1225
11.19
98.2
19.0 66.2
130
1251
11.43
100.1
20.0
68
133 1278
11.67
102.1
21.0
69.8
136
1305
11.91
104.0
22.0
71.6
139
1332
12.15
105.4
23.0
73.4
142
1359
12.39
107.9
24.0
75.2
145
1386
12.64
109.8
25.0
77
147
1413
12.88
111.8
26.0
78.8
150
1440
13.13
113.8
27.0
80.6
153
1467
13.37
115.7
28.0
82.4
156
1494
13.62
--
29.0
84.2
159
1522
13.87
--
30.0
86
162
1549
14.12
--
31.0
87.8
165
1581
14.37
--
32.0
89.6
168
1609
14.62
--
33.0
91.4
171
1638
14.88
--
34.0
93.2
174
1667
15.13
--
35.0
95
177
1696
15.39
--