For battery replacement, refer to
. Make sure that the cover is
tightly closed to maintain the IP67 enclosure rating.
Figure 12 Battery replacement
Troubleshooting
Refer to the following table for common problem messages or
symptoms, possible causes and corrective actions.
Error/Warning
Description
Solution
Measurement out
of range.
• Insert the probe into the appropriate
standard and read again.
• Clean the probe and the membrane.
• Make sure that there are no air
bubbles in the membrane.
• Disconnect the probe and then
connect the probe again.
• Connect a different probe to verify if
problem is with probe or meter.
Battery is low.
Insert new batteries.
E1
Unstable reading
during a stability
measurement or
during calibration.
Make sure that the probe is properly
immersed in the sample.
E2
Probe current:
exceeded upper
limit during
measurement ≥
250 nA.
• Clean the probe and the membrane.
• Make sure that there are no air
bubbles in the membrane.
• Disconnect the probe and then
connect the probe again.
• Connect a different probe to verify if
problem is with probe or meter.
E2
pH probes
sensitivity out of
range (accepted
values 70–105%).
• Make sure that the correct probe is
connected to the meter.
• Connect a new probe.
16
English
Summary of Contents for sensION+ EC5 DL
Page 2: ...English 3 Fran ais 20 Espa ol 40 Portugu s 58 77 93 111 128 2...
Page 80: ...1 IP67 1 1 2 2 2 3 1 5 2 6 3 7 4 80...
Page 81: ...3 4 PH1 DL 1 4 2 5 3 6 pH ORP 5 DO6 DL 1 4 2 5 3 6 DO 81...
Page 82: ...6 MM110 DL 1 5 2 6 ORP 3 7 pH 4 7 MM150 DL 1 6 2 pH 7 TDS ORP 3 8 4 9 5 82...
Page 83: ...8 MM156 DL 1 6 2 pH 7 DO 3 8 4 9 5 9 EC5 DL 1 5 2 6 TDS 3 7 4 83...
Page 88: ...10 portCom USB RF 1 USB RF USB USB RF 2 portCom 3 Snd portCom 11 4 5 6 txt csv 11 88...
Page 89: ...12 IP67 12 E1 E2 250 nA E2 pH 70 105 E2 30 89...
Page 96: ...1 IP67 1 1 2 2 2 3 1 5 2 6 3 7 4 96...
Page 97: ...3 4 PH1 DL 1 4 hh mm ss 2 5 3 6 pH ORP 5 DO6 DL 1 4 hh mm ss 2 5 3 6 DO 97...
Page 98: ...6 MM110 DL 1 5 2 6 ORP 3 7 pH 4 hh mm ss 7 MM150 DL 1 6 2 pH 7 TDS ORP 3 8 4 9 5 hh mm ss 98...
Page 99: ...8 MM156 DL 1 6 2 pH 7 DO 3 8 4 9 5 hh mm ss 9 EC5 DL 1 5 hh mm ss 2 6 TDS 3 7 4 1 99...
Page 100: ...5 MSDS SDS EMI 20 cm 20 cm pH pH ORP 5 10 5 15 45 15 6 100...
Page 105: ...3 2 Snd portCom 11 4 Save 5 6 txt csv Save 11 3 105...
Page 106: ...12 IP67 12 E1 E2 250 nA E2 pH 70 105 106...
Page 107: ...E2 30 E3 0 05 cm 1 E3 58 mV E4 E4 TC 0 TC E5 E6 ErA MM156 50 g L TC 107...
Page 110: ...3 mS cm g L NaCl 41 2 25 4 48 9 30 6 56 3 35 8 62 2 40 0 69 4 45 3 75 7 50 0 110...
Page 115: ...3 4 PH1 DL 1 4 hh mm ss 2 5 3 6 pH ORP 5 DO6 DL 1 4 hh mm ss 2 5 3 6 DO 115...
Page 116: ...6 MM110 DL 1 5 2 6 ORP 3 7 pH 4 hh mm ss 7 MM150 DL 1 6 2 pH 7 TDS ORP 3 8 4 9 5 hh mm ss 116...
Page 117: ...8 MM156 DL 1 6 2 pH 7 DO 3 8 4 9 5 hh mm ss 9 EC5 DL 1 5 hh mm ss 2 6 TDS 3 7 4 117...
Page 118: ...5 MSDS SDS EMI 20 cm 7 9 20cm 7 9 in pH pH ORP 5 10 5 15 45 15 6 118...
Page 123: ...11 123...
Page 124: ...12 IP67 12 E1 E2 250nA E2 pH 70 105 124...
Page 125: ...E2 30 E3 0 05cm 1 E3 58mV E4 E4 TC 0 TC E5 E6 ErA MM156 50 g L TC E7 E8 E9 USB RF portCom 125...
Page 131: ...1 IP67 1 1 2 2 2 3 1 DISPLAY LIGHT 5 2 DATALOG 6 ON OFF 3 7 CALIBRATION 4 MEASUREMENT 131...
Page 132: ...3 Measure Calibration Standby 4 PH1 DL 1 4 2 5 3 6 pH ORP 5 DO6 DL 1 4 2 5 3 6 DO 132...
Page 133: ...6 MM110 DL 1 5 2 6 ORP 3 7 pH 4 7 MM150 DL 1 6 2 pH 7 TDS ORP 3 8 4 9 5 133...
Page 134: ...8 MM156 DL 1 6 2 pH 7 DO 3 8 4 9 5 9 EC5 DL 1 5 2 6 TDS 3 7 4 datalog 134...
Page 135: ...5 MSDS SDS EMI 20 7 9 20 7 9 pH pH ORP 5 10 5 15 45 15 6 135...
Page 140: ...11 140...
Page 141: ...12 IP67 12 E1 E2 250 nA E2 pH E2 30 141...
Page 145: ......