20
Installation
5.
Move the tension lever to the left until it clicks into place (
). Make sure the
tension lever is in the 12 o'clock position.
Note:
Move the lever to the right to remove tension.
6.
Store unused cartridges.
3.2.4 Reagent pump initial test
To test the reagent RP-150 pump:
1.
Make sure that the Omnion software is not active at this time. Press the
NORMAL
RUN
button. Make sure the pump is running at a normal speed of 35. The NORMAL
LED will be illuminated. If the display shows a different speed, press the
ARROW
button to change the setting to 35.
2.
Press the
MIN
button. The MIN LED will be illuminated. The display will show a speed
of 4.
3.
Hold down the
MAX
button until the MAX LED is illuminated. The display will show a
speed of 999. The pump will go back to the original speed when the
MAX
button
is released.
4.
To stop the pump, press the
MANUAL RUN/ STOP
button. No LEDs will be illuminated,
but the display will show “35” (the pump speed).
5.
Press the
MANUAL RUN/ STOP
button again. The MANUAL LED will be illuminated.
The pump will be running at normal speed of 35.
6.
Press the
MANUAL RUN/ STOP
button to stop the pump. To repeat any of the steps
above, turn off the pump and start with step 1.
7.
If any of the above steps fail, contact Technical Support or a local
sales representative.
Note:
Contact Technical Support for queries regarding the installation and testing of a different
peristaltic pump model.
Figure 18 Tension lever setup
1
One click back
Summary of Contents for QC8500 Series 2
Page 2: ......
Page 6: ...4 Table of Contents ...
Page 10: ...8 Specifications ...
Page 30: ...28 Installation ...
Page 52: ...50 Operation ...
Page 84: ...82 Troubleshooting ...
Page 88: ...86 Replacement Parts and Accessories ...
Page 90: ...88 ...
Page 92: ...90 Warranty ...
Page 94: ...92 Certification ...
Page 102: ...100 Ion Chromatography ...