Specification
Details
Outputs
Analog output (temperature, conductivity/resistivity): 2 × 0/4-20 mA (linear, bilinear,
logarithmic) ± 0.1 mA
Alarms: 2 × thresholds or limits according to USP
Certifications
EN 61326-1: 2006; EN 61010-1: 2010
Sensor
Specification
Details
Sensor body material
Black PSU
Conductivity electrodes, internal and external
Stainless steel 316L
Cell constant K
0.01 (cm
-1
)
Conductivity range
0.01—200 μS.cm
-1
; Resistivity range: 5k Ω.cm—100 MΩ.cm
Maximum pressure
10 bar
Maximum temperature
125 °C (257 °F)
Accuracy
< 2%
Temperature response
< 30 seconds
Insulator
PSU
Connector
Glass polyester (IP65)
Controller
Specification
Details
Component description
Microprocessor-controlled and menu-driven controller that operates the sensor and
displays measured values.
Operating temperature
-20 to 60 ºC (-4 to 140 ºF); 95% relative humidity, non-condensing with sensor load
<7 W; -20 to 50 ºC (-4 to 104 ºF) with sensor load <28 W
Storage temperature
-20 to 70 ºC (-4 to 158 ºF); 95% relative humidity, non-condensing
Enclosure
1
NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish
Power requirements
AC powered controller:
100-240 VAC ±10%, 50/60 Hz; Power 50 VA with 7 W
sensor/network module load, 100 VA with 28 W sensor/network module load
(optional Modbus, RS232/RS485, Profibus DPV1 or HART network connection).
24 VDC powered controller:
24 VDC—15%, + 20%; Power 15 W with 7 W
sensor/network module load, 40 W with 28 W sensor/network module load (optional
Modbus, RS232/RS485, Profibus DPV1 or HART network connection).
Altitude requirements
Standard 2000 m (6562 ft) ASL (Above Sea Level)
Pollution
degree/Installation
category
Polution Degree 2; Installation Category II
Outputs
Two analog (0-20 mA or 4-20 mA) outputs. Each analog output can be assigned to
represent a measured parameter such as pH, temperature, flow or calculated
values. Optional module supplies three additional analog outputs (5 total).
1
Units that have the Underwriters Laboratories (UL) certification are intended for indoor use only
and do not have a NEMA 4X/IP66 rating.
4
English
Summary of Contents for Polymetron 9526
Page 130: ...1 2 3 4 5 2 DN8 DN6 DN6 2 IN OUT 100 s cm CO2 1 2 S cm KCI 10 S cm 130...
Page 134: ...3 0 35 2 mm2 AWG 22 14 105 4 7 1 131 4 1 5 9 2 6 10 3 7 4 8 1 6 4 6 2 1 4 134...
Page 136: ...5 1 5 Back 2 6 Menu Settings Menu 3 HOME Main Measurement 7 4 ENTER SD 6 136...
Page 206: ...EMC 206...
Page 207: ...EMC A A IECS 003 A FCC 15 15 FCC 1 2 15 FCC 1 2 3 4 5 207...
Page 208: ...2 DN8 DN6 DN6 2 100 CO2 1 2 10 KCl 208...
Page 212: ...30 2 1 D1 2 D2 OUT 20 60 30 212...
Page 213: ...3 0 35 2 2 AWG 22 14 105 C 4 7 1 209 213...
Page 215: ...Startup 203 1 2 3 4 5 5 5 1 5 BACK 2 SD 6 MENU 3 HOME 7 4 ENTER SD 215...
Page 245: ......