Table of contents
on page 3
on page 16
on page 16
on page 8
Replacement parts and accessories
on page 16
on page 16
Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification
Details
Operating temperature
–10 to 50 °C (14 to 122 °F)
Storage temperature
–5 to 40 °C (23 to 104 °F)
Relative humidity
20 to 80%, non-condensing
Altitude
2000 m (6562 ft) maximum
Overvoltage category
2
Pollution degree
2
Power requirements
24 VAC, 110–240 VAC, 50–60 Hz, 10 VA
CE compliance
EN61326-1: EMC Directive
EN61010-1: LVD Directive
Warranty
1 year (EU: 2 years)
Table 2 Specifications—Air or water electrode cleaning system
Specification
Details
Dimensions (W x H x D) 55 x 100 x 40 mm (9.8 x 13.8 x 5.9 in.)
Weight
Approximately 0.4 kg (9.9 lbs)
Cleaning system
Pressure: 0.5 to 10 bars (7.25 to 145 psi); temperature 70 °C (158 °F) maximum; flow:
400 to 4000 L/h
Environmental rating
IP 65
Materials
Enclosure material: Brass (MS); seal material: Fluoroelastomer (FKM), etylene
propylene diene monomer rubber (EPDM) and nitrile rubber (NBR); tubing:
Polyethylene (PE)
Table 3 Specifications—Chemical electrode cleaning system
Specification
Details
Dimensions (W x H x D) 250 x 350 x 150 mm (9.8 x 13.8 x 5.9 in.)
Weight
Approximately 4.5 kg (9.9 lbs)
Cleaning system
Chemical reagent: air supply pressure 4 to 8 bar (58 to 116 psi); volume of reagent
supplied in each cycle: ± 10 mL
Suction height
2 m (6.5 ft) maximum
English
3