English
120
Installation
Please only perform the following work steps as a
pair:
1.
Plug in power plug, device starts in the [+SERVICE] menu.
2.
Open housing door and remove transparent safety panel.
3.
Pull out power plug again and place the measurement device on
the refrigerator (first set thermostat knob to 5–6), then route the
solenoid valve control cable (at the back of the refrigerator) from
behind into the housing of the measurement device (
4.
Unscrew the photometer baseplate and lift up forwards; connect
control cable to the fan plate, which is located on the housing back
).
Figure 3 Refrigerator valve connection
Figure 4 Fan plate
Connections of the feeds and drains
On the left housing side is the sample feed hose (approximately 1.2 m
long), which under no circumstances may be shortened or lengthened. It
is specially designed for the sample feeding through SIGMATAX 2.
Above this is a PG screw connection with outside wall connector for the
WA R N I N G
Electrical dangers and fire hazard. Use only the supplied power cable
Only qualified experts may perform the tasks described in this section of the
manual, in compliance with all locally applicable safety regulations.
N O T I C E
Use only earthed sockets for the connection of this device to the power supply.
If you are not sure if the sockets are earthed, have this checked by a qualified
electrician.
The power plug serves in addition to the power supply to isolate the device
quickly from the mains where necessary.
The entire measurement system has two power plugs (measurement device
and refrigerator). During the disconnection from the mains it must be made sure
that the correct power plug is pulled (e.g. labeling of the sockets).
This is recommended for long-term disuse and can prevent potential dangers in
the event of a fault.
Therefore make sure that the sockets to which the device is connected are easy
to reach by each user at all times.
N O T I C E
As the refrigerator itself is not provided with a fuse, it may only be connected to a
10 A fused CEE socket.
ST
O
P
1
2
3
4
5
6
7
Set thermostat knob
at the top of the
refrigerator to 5–6.
Control cable
Photometer-
baseplate
Control cable connection
Summary of Contents for PHOSPHAX sigma High Range
Page 5: ...5 II 2 230 V AC 10 50 Hz 145 VA x 540 mm x 600 mm x 400 mm 22 kg...
Page 7: ...7 12 2005 2002 96 Hach Lange Hach Lange...
Page 9: ...9 1 SERVICE 2 3 5 6 3 4 4 3 CEE 10 A STO P 1 2 3 4 5 6 7 5 6...
Page 10: ...10 4 1 2 m SIGMATAX 2 PG LZP 573 5 2 m 5 PG 6 9 mm 1 4 m T 2 5 A L 250 V...
Page 11: ...11 6 SIGMATAX 2 SIGMATAX 2 1 2...
Page 15: ...15 15 PHOSPHAX sigma 1 2 m 5 100 C SERVICE C D 9 48...
Page 16: ...16 3 A C D 6 6 1 2 a SERVICE 18 SERVICE 3 6 Hach Lange 10 30...
Page 17: ...17 Q1 Q2 QA A A PHOSPHAX sigma 11 10 C 5 6 48 6 7 PHOSPHAX...
Page 29: ...29 PHOSPHAX sigma 12 6 O A C D 3 3 18 24...
Page 32: ...32...
Page 114: ...Deutsch 114...
Page 226: ...Fran ais 226...
Page 282: ...Italiano 282...
Page 370: ...Rom n 370...
Page 452: ...T rk e 452...
Page 456: ...456 2005 8 12 2002 96 EC Hach Lange Hach Lange...
Page 463: ...463 100 C SERVICE C D 9 3 A C D 6 6 1 2 48 Pump cassette Pump roller...
Page 464: ...464 SERVICE 466 SERVICE 3 6 Hach Lange 10 Q1 Q2 QA A A PHOSPHAX sigma 30 48...
Page 465: ...465 11 STATUS 10 C 5 6 6 7 PHOSPHAX...
Page 476: ...476 PHOSPHAX sigma 12 6 O A C D 3 3 18 24...
Page 477: ...477 Status STATUS SIGMATAX SIGMATAX STATUS SIGMATAX SIGMATAX SIGMATAX STATUS...
Page 479: ...480...