Installation
3.4.2 Remove the shipping transport locks
Prior to system start-up, the shipping transport locks must be
removed from the sc analyzer.
CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.
Important note:
The cuvette insulation for the measuring unit is
NOT a transport lock. Do NOT remove the cover on the measuring
unit.
1.
Open the enclosure door and secure with the door hook.
2.
Remove the transport lock on the analyzer panel (
Note:
If the instrument is operated with a Filter Probe sc, it is equipped
with an internal compressor.
3.
Remove the cable tie and pull out the compressor transport
lock to the left (
).
Note:
Keep the transport locks for transporting and storage.
Figure 9 Remove the analyzer panel transport locks
1
Transport lock