Hach PHC28101 User Manual Download Page 6

6.

Push 

Read

. The display will show "Stabilizing" and a progress bar as the reading

stabilizes. The display shows the value of the check standard and either Check
Standard Passed or Check Standard Failed.

7.

If the display shows 

Check Standard Passed

, the check standard measurement is

within the accepted limits set by the administrative user. Select 

Done

 to continue with

the sample measurement.

8.

If the display shows 

Check Standard Failed

, the measurement is outside of

accepted limits set by the administrative user and a recalibration is recommended. If
the acceptance criteria is set to Cal Expires on Failure: Yes, the display shows the
calibration icon and a question mark until the probe is recalibrated. To correct the
probe calibration and status indicator, calibrate the probe (refer to 

Calibration

on page 3).

Advanced operation

Parameter-specific settings can be changed through the Full Access Options menu.
Details about menu navigation, available options and how to change them are given in
the screens, tables and procedures throughout this section.

The settings that can be changed are shown in 

Table 1

.

Table 1  Parameter-specific settings

Setting

Options

Measurement Options

Resolution

Upper and lower range limits

Calibration Options

Buffer set

Buffer set values (if Custom Buffer Set selected)

Reminder

Minimum Cal points

Slope limit

Check Standard Options

Standard (temperature compensated buffer or custom at 25 °C)

Reminder

Acceptance criteria

Standard value at 25 °C (if custom standard selected)

Units

pH

mV

6

Summary of Contents for PHC28101

Page 1: ...combination probe with refillable double junction reference and built in temperature sensor pH range pH 0 to 14 pH resolution User selectable stabilization time and resolution Fast 0 1 Fast 0 01 Medi...

Page 2: ...viscous storage filling solution completely Blot dry with a lint free cloth 4 Remove the protective tape from the filling hole before initial use refer to Figure 1 on page 2 Dispose of the protective...

Page 3: ...ne with a pH above and one with a pH below the expected sample pH For a one point calibration select the buffer nearest to the expected sample pH The calibration is recorded in the probe and the data...

Page 4: ...rect service life time stamp Set the date and time in the meter before the probe is attached If complete traceability is necessary enter a sample ID and operator ID before measurement Refer to the HQd...

Page 5: ...eat steps 2 4 for additional measurements 7 When measurements are done store the probe refer to Storage on page 10 Run a check standard The run check standard feature validates instrument performance...

Page 6: ...question mark until the probe is recalibrated To correct the probe calibration and status indicator calibrate the probe refer to Calibration on page 3 Advanced operation Parameter specific settings c...

Page 7: ...elect Measurement Options and update the settings Option Description Resolution Sets the resolution 0 1pH Fast 0 01pH Fast default 0 01pH Medium 0 01pH Slow or 0 001pH Slow The resolution affects the...

Page 8: ...ration points necessary before a calibration can be completed 1 default 2 or 3 Slope Limit Sets the slope limit 1 to 10 acceptable slope criteria default 5 The slope must fall within set limits for su...

Page 9: ...ed when this option is selected Change check standard options 1 Make sure a probe is connected to the meter 2 Push and select PHC281 Settings 3 Select Modify Current Settings 4 Select Check Standards...

Page 10: ...ute in deionized water 4 Soak the probe in pH 4 buffer for up to 20 minutes then rinse with deionized water 5 Blot dry with a lint free cloth For fats grease and oils 1 Soak the glass bulb in a warm d...

Page 11: ...rer s website Refer to the HQd Series meter manual for specific instructions for the meter model HQd meter does not support IntelliCAL probe Contact a Technical Support Representative Connect a probe...

Page 12: ...ts by changing the PHC281 calibration settings and method 2 Recalibrate the probe 3 Run a check standard to check the performance of the probe Drifting Inaccurate readings Contaminated glass bulb Clea...

Page 13: ...the current method Create a new method with an expanded range Temperature out of range Temperature value is outside of range Make sure that the sample temperature is within the range of the probe Mak...

Page 14: ...hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compois 1222 V...

Reviews: